10-16
IM 760401-01E
10.7 Commands (before the IEEE 488.2 Standard)
For a detailed description of each command, refer to section 13.1.
Command
Description
Wiring system
WR m (WiRing)
sets wiring system
Voltage range
RV m (Range Voltage)
sets voltage range
AV m (Auto Voltage range)
sets voltage auto range
Current range
RA m (Range current(A))
sets current range
AA m (Auto current(A) range)
sets current auto range
SA m (Sensor Ampere)
sets external sensor scalling value
Measurement mode MN m (MeaN)
sets RMS, VOLTAGE MEAN, or DC
Measurement
SN m (SyNc source)
sets the measurement synchronization source
synchronization
source
Line filter
LF m (Line Filter)
sets line filter ON/OFF
Frequency filter
FL m (FiLter)
sets frequency filter ON/OFF
Update rate
SI m (Sample Interval)
sets display update rate
Hold
HD m (sampling HolD)
holds display and output data
Trigger
E or ST or <GET>
trigger
Display
DA m (Display A function)
selects function to be displayed on display A
DB m (Display B function)
selects function to be displayed on display B
DC m (Display C function)
selects function to be displayed on display C
EA m (Element display A)
selects element to be displayed on display A
EB m (Element display B)
selects element to be displayed on display B
EC m (Element display C)
selects element to be displayed on display C
Scaling
SC m (SCaling)
sets scaling ON/OFF
KV m (K*Amplre)
sets the scaling value
KA m (K*Wattage)
KW m (K*Voltage)
Averaging
AG m (AveraGing)
sets averaging ON/OFF
AT m (Averaging Type)
selects exponential averaging or moving
AC m (Averaging Coefficient)
averaging sets attenuation constant or
averaging number
Crest factor
CF m (Crest Factor)
Sets the crest factor
MAX hold
KH m (peaK Hold)
sets MAX hold
MATH
MT m (MaThematics)
sets computing equation
Number of displayed DS m (Display reSolution)
sets the number of displayed digits
digits
Display range
DR (Display Range)
displays the current range
Key lock
KL m (Key Lock)
sets key lock ON/OFF
Zero-level
ZC (Zero Calibration)
executes zero-level compensation
compensation
Integration
IS (Integrate Start )
starts integration
IP (Integrate stoP)
stops integration
IR (Integrate Reset)
resets integration
IC m (Integrate Continuous)
sets integration mode
TM m1,m2,m3 (integrate TiMer)sets integration preset time
Data storage
SO (Store On)
starts storage
SR m1,m2,m3 (Store inteRval) sets storage interval
Data recalling
RO m (Recall On)
starts recalling
RR m1,m2,m3 (recall inteRval) sets recalling interval
Setup parameters SL m (panel Setting Load)
recalling setup parameters
SS m (panel Setting Save)
storing setup parameters
RC (Reset Command)
initialize setup parameters
Communication
CM m (Communication coMmand) sets command group to be used
commands
OD (Output Data)
requests output of measured data
OF m1,m2,m3 (Output Function) sets output items
OFD m (Output Function Default) sets default output items
OS (Output panal Setting)
requests output of setting parameters
OE (Output Error code)
requests output of error code
H m (Header)
sets output data header
DL m (DeLimiter)
sets output data delimiter
IM m (Interrupt Mask)
sets status byte interrupt mask
Summary of Contents for WT210
Page 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Page 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Page 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Page 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Page 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...