10-12
IM 760401-01E
Explanation
Setting the Mode
Refer to page 10-1 for details.
Setting the Address
A particular address is assigned to each device connected to the GP-IB interface so
that each device can be recognized by every device. Therefore, an address must be
assigned to this instrument when it is connected to a personal computer.
Address setting range: 0 to 30
The initial value is “1.” The address is retained even if the instrument is initialized.
Talk-only Function
This function only allows the instrument to send data to other devices. Data can be sent even
when talk-only is OFF. In talk-only mode, the instrument cannot be controlled by a controller.
Interval
In case of the talk-only mode, this setting specifies the interval to output data.
Setting range: 00.00.00 (0 hr 00 min 00 sec) to 99.59.59 (99 hrs 59 min 59 sec)
Initial value: 00.00.00
When set to 00.00.00, the interval is equal to the display update rate. In addition,
when the specified interval is shorter that the display update rate, the data is output
using the display update rate.
Terminator
When this instrument is used as a listener
Use “CR+LF,” “LF,” or “EOI” as the receiving terminator.
When this instrument is used as a talker
The sending terminator is set using the DL command. The initial setting is “CR+LF+EOI.”
Note
It is not possible for this instrument to receive data if the “CR” terminator is sent from the controller. It
is also not possible to set “CR” as the terminator which is to be sent from this instrument.
10.5 Setting the Address and Mode
Summary of Contents for WT210
Page 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Page 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Page 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Page 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Page 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...