9-5
IM 760401-01E
External In/Output Function
9
Setting the Preset integration Time
1.
SHIFT
RESET
INTEG SET
ENTER
3.
Select the preset integration time
( Display C )
2.
Hour
( Display A )
ENTER
6.
4. sets the value
5. moves digits
SHIFT
Minute
( Display B )
ENTER
9.
7.
8.
Same as steps
4 and 5 above
Second
( Display C )
ENTER
12.
End of setting
10.
11.
Same as steps
4 and 5 above
Explanation
D/A Output
Voltage, current, active power, apparent power, reactive power, power factor, phase
angle, harmonic measurement data and integrated data values will be output as a 5 V
FS analog voltage. The number of items which can be output (number of output
channels) depends on the installed options.
Default Setting of the Output Format (D/A)
The default items which will be output can be selected as follows.
• dFLt-n (normal measurement values are set as default)
Select this when you want to output normal measurement values. Which items are
output to which channel is described below.
Option
Model
ch1
ch2
ch3
ch4
ch5
ch6
ch7
ch8
ch9
ch10
ch11
ch12
/DA4
760401
/DA12
/CMP
760502
760503
760401
760502
760503
V
A
W
Hz
V1
-
V3
Σ
V
A1
-
A3
Σ
A
W1
-
W3
Σ
W
V1
V2
V3
Σ
V
A1
A2
A3
Σ
A
W1
W2
W3
Σ
W
V
A
W
Hz
These
channels
cannot be
set.
These channels cannot be set.
*1
*1
*1
*2
If either one of the display function, V Hz or A Hz, is turned on, the frequency of the
corresponding function is output. If both display functions are OFF, the voltage frequency
of the element assigned to display C is output.
The number corresponds to input element 1, 2, or 3.
*2
*2
V1
A1
W1
Hz
*1
V1
A1
W1
Hz
*1
*2
*2
Output
channel
9.3 D/A Output (Option)
Summary of Contents for WT210
Page 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Page 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Page 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Page 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Page 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...