1-4
IM WT1801-01EN
Harmonic Measurement Function Orders
The harmonic orders that you can specify are indicated below.
Input Element Harmonic Measurement Functions
Measurement
Function
Characters or Numbers in Parentheses
Total
0 (DC)
1
k
U( )
Yes
Yes
Yes
2 to 500
I( )
Yes
Yes
Yes
2 to 500
P( )
Yes
Yes
Yes
2 to 500
S( )
Yes
Yes
Yes
2 to 500
Q( )
Yes
Fixed at 0
Yes
2 to 500
λ
( )
Yes
Yes
Yes
2 to 500
Φ
( )
Yes
No
Yes
2 to 500
Φ
U( )
No
No
No
2 to 500
Φ
I( )
No
No
No
2 to 500
Z( )
No
Yes
Yes
2 to 100
Rs( )
No
Yes
Yes
2 to 100
Xs( )
No
Yes
Yes
2 to 100
Rp( )
No
Yes
Yes
2 to 100
Xp( )
No
Yes
Yes
2 to 100
Uhdf( )
No
Yes
Yes
2 to 500
Ihdf( )
No
Yes
Yes
2 to 500
Phdf( )
No
Yes
Yes
2 to 500
Uthd
Yes
No
No
No
Ithd
Yes
No
No
No
Pthd
Yes
No
No
No
Uthf
Yes
No
No
No
Ithf
Yes
No
No
No
Utif
Yes
No
No
No
Itif
Yes
No
No
No
hvf
Yes
No
No
No
hcf
Yes
No
No
No
K-factor
Yes
No
No
No
Functions with parentheses will produce different values depending on which of the following is contained in their
parentheses.
• Total: Total value (The total value of all harmonic components from the minimum order to N.
*
For information
about how the value is determined, see appendix 1 in the getting started guide, IM WT1801-03EN.)
• 0(DC): DC value
• 1:
Fundamental harmonic value
• k:
The value of any order from 2 to N.
*
* N is the maximum measurable order. The maximum measurable harmonic order is the smallest of the three
orders listed below.
• The specified maximum measurable harmonic order
• The value determined automatically according to the PLL source frequency (see section 6.4 in the getting
started guide, IM WT1801-03EN)
• When the data update interval is 50 ms, the maximum measurable harmonic order is 100.
1 Items That This Instrument Can Measure