12-2
IM WT18O1E-02EN
12.2 Setting the Element and Wiring Unit to Display
and the Zoom Factor
This section explains the following vector display settings. This feature is available on models with the
/G5 or /G6 option.
• Vector number
• Element and wiring unit
• Zoom factor
►
“Display Items (ITEM)
—Vector
” in the features guide
1.
Follow the procedure in section 12.1 to select Vector on the Others menu.
Vector Items Menu
2.
Press
ITEM
to display the following menu.
• If the setup parameter list is being displayed (the INPUT INFO key is illuminated), the Info Form menu
may be displayed. If this happens, press
ITEM
again.
• If setup parameter list is being displayed, the vector that you have set to vector number 1 is displayed in
the bottom half of the screen.
Select the vector number that you want to set (1, 2).
Set the element and wiring unit (Element 1 to Element 6, ΣA to ΣC).
Set the zoom factor (0.100 to 100,000).
• Set the zoom factor of fundamental wave U (1) or I (1). The value that indicates the size of
the vector display
’
s peripheral circle changes according to the zoom factor, and the size of
the vectors that indicate U (1) and I (1) change accordingly as well.
• If you press this soft key to select both U Mag and I Mag, you can link the zoom factors of
both settings and change them at the same time.