User-Defined Events (User Defined Event)
User-defined events can be used to trigger data storage and automatic printing (option). You can define up to
eight user-defined events.
User-Defined Event Number (Event No.)
Select a user-defined event number from 1 to 8.
Turning User-Defined Events On and Off
Select whether to enable user-defined events.
User-Defined Event Name (Event Name)
• Number of characters: Up to eight
• Usable characters: Spaces and all characters that are displayed on the keyboard
Indication of User-Defined Event Occurrence (TRUE/FALSE)
Set the character strings to show when a user-defined event occurs and when it does not occur.
• Number of characters: Up to six
• Usable characters:
Spaces and all characters that are displayed on the keyboard
Judgment Condition Specification Method (Expression)
Select the method for specifying judgment conditions from the following options.
• Range (Range): Specify judgment conditions using ranges of measurement function values or differences
from a reference value.
• Condition (Condition):
Specify judgment conditions using user-defined events.
Range (Range)
Specify the judgment condition here when you have set the judgment condition specification method to Range.
• Function (Function)
You can select any of the measurement function types listed under “Items That This Instrument Can
Measure.”
• Element (Element/Σ)
You can select the element/wiring unit from the options below. The available options vary depending on the
installed elements.
Element1, Element2, Element3, Element4, Element5, Element6, ΣA, ΣB, and ΣC
• Order (Order; option)
This is the same as setting the harmonic order in the 4-, 8-, and 16-value displays.
• Judgment Condition
Select the method for comparing the measured value and the reference value from the following options.
OFF, <, <=, =, >, >=, != (not equal)
• Reference Value
Select a number from –9.9999 T to 9.9999 T.
8 Computation