5-55
IM 760151-17E
Commands
5
[:INPut]:WIRing
Function
Sets the wiring system or queries the current
setting.
Syntax
[:INPut]:WIRing {(P1W2|P1W3|P3W3|
P3W4|V3A3)[,(P1W2|P1W3|P3W3|P3W4|
V3A3|NONE)][,(P1W2|P1W3|P3W3|P3W4|
V3A3|NONE)]}
[:INPut]:WIRing?
P1W2 = Single-phase, two-wire system
P1W3 = Single-phase, three-wire system
P3W3 = Three-phase, three-wire system
P3W4 = Three-phase, four-wire system
V3A3 = Three voltage, three current system
NONE = No wiring
Example
:INPUT:WIRING P1W2,P1W2,P1W2
:INPUT:WIRING? -> :INPUT:
WIRING P1W2,P1W2,P1W2
Description • Set the wiring systems in the order
Σ
A,
Σ
B,
and
Σ
C.
• If the combination does not allow setting of
Σ
B or
Σ
C, it can be omitted.
• Certain combinations of wiring systems are
not selectable depending on the model type.
For the combinations of wiring systems, see the
WT1600FC User’s Manual (IM760151-01E).
• On models with a single power measurement
element,
Σ
A is fixed to P1W2.
Σ
B and
Σ
C
cannot be specified.
• This command is invalid on models that are
only equipped with impedance measurement
elements, since the mode is fixed to
impedance measurement.
5.10 INPut Group