5-43
IM 760151-17E
Commands
5
:IMPedance[:STATe]
Function
Turns ON/OFF the impedance measurement
mode or queries the current setting.
Syntax
:IMPedance[:STATe] {<Boolean>}
:IMPedance:STATe?
Example
:IMPEDANCE:STATE ON
:IMPEDANCE:STATE? ->
:IMPEDANCE:STATE 1
:IMPedance:SUPerpose?
Function
Queries all settings related to the load current
for impedance measurements.
Syntax
:IMPedance:SUPerpose?
Example
:IMPEDANCE:SUPERPOSE? ->
:IMPEDANCE:SUPERPOSE:OUTPUT:
STATE 0;TYPE DC;:IMPEDANCE:
SUPERPOSE:FREQUENCY:RANGE HZ;
VALUE 1.0002;:IMPEDANCE:SUPERPOSE:
OFFSET 0.000;AMPLITUDE 0.000;
DETAILE:RATIO 1.000;LIMIT 1000.000,
-1000.000;HOLD NORMAL;WAVEFORM SINE
:IMPedance:SUPerpose:AMPLitude
Function
Sets the amplitude of the load current for
impedance measurements or queries the
current setting.
Syntax
:IMPedance:SUPerpose:
AMPLitude {<NRf>}
:IMPedance:SUPerpose:AMPLitude?
<NRf>
= 0.000 to 2000.000(App)
Example
:IMPEDANCE:SUPERPOSE:AMPLITUDE 0
:IMPEDANCE:SUPERPOSE:AMPLITUDE? ->
:IMPEDANCE:SUPERPOSE:
AMPLITUDE 0.000
:IMPedance:SUPerpose:DETaile?
Function
Queries all settings related to the detailed
settings of the load current for impedance
measurements.
Syntax
:IMPedance:SUPerpose:DETaile?
Example
:IMPEDANCE:SUPERPOSE:DETAILE? ->
:IMPEDANCE:SUPERPOSE:DETAILE:
RATIO 1.000;LIMIT 1000.000,
-1000.000;HOLD NORMAL;WAVEFORM SINE
:IMPedance:SUPerpose:DETaile:HOLD
Function
Sets the action taken by the WT1600FC
(handling of the control signal to the impedance
measurement electronic load device) when hold
is activated or queries the current setting.
Syntax
:IMPedance:SUPerpose:DETaile:
HOLD {NORMal|OFF|AC_OFF}
:IMPedance:SUPerpose:DETaile:HOLD?
Example
:IMPEDANCE:SUPERPOSE:DETAILE:
HOLD NORMAL
:IMPEDANCE:SUPERPOSE:DETAILE:HOLD?
-> :IMPEDANCE:SUPERPOSE:DETAILE:
HOLD NORMAL
:IMPedance:SUPerpose:DETaile:LIMit
Function
Sets the range of the load current for
impedance measurements or queries the
current setting.
Syntax
:IMPedance:SUPerpose:DETaile:
LIMit {<NRf>,<NRf>}
:IMPedance:SUPerpose:DETaile:LIMit?
<NRf>
= –1000.000 to 1000.000(A)
Example
:IMPEDANCE:SUPERPOSE:DETAILE:
LIMIT 1000,-1000
:IMPEDANCE:SUPERPOSE:DETAILE:LIMIT?
-> :IMPEDANCE:SUPERPOSE:DETAILE:
LIMIT 1000.000,-1000.000
Description Set the upper limit and then the lower limit.
:IMPedance:SUPerpose:DETaile:RATio
Function
Sets the current value per volt of the control
signal to the impedance measurement
electronic load device or queries the current
setting.
Syntax
:IMPedance:SUPerpose:DETaile:
RATio {<NRf>}
:IMPedance:SUPerpose:DETaile:RATio?
<NRf>
= 0.001 to 1000.000
Example
:IMPEDANCE:SUPERPOSE:DETAILE:
RATIO 1
:IMPEDANCE:SUPERPOSE:DETAILE:RATIO?
-> :IMPEDANCE:SUPERPOSE:DETAILE:
RATIO 1.000
:IMPedance:SUPerpose:DETaile:WAVeform
Function
Sets the waveform of the load current for
impedance measurements or queries the
current setting.
Syntax
:IMPedance:SUPerpose:DETaile:
WAVeform {SINE|PATTern}
:IMPedance:SUPerpose:DETaile:
WAVeform?
Example
:IMPEDANCE:SUPERPOSE:DETAILE:
WAVEFORM SINE
:IMPEDANCE:SUPERPOSE:DETAILE:
WAVEFORM? -> :IMPEDANCE:SUPERPOSE:
DETAILE: WAVEFORM SINE
:IMPedance:SUPerpose:FREQuency?
Function
Queries all settings related to the frequency of
the load current for impedance measurements.
Syntax
:IMPedance:SUPerpose:FREQuency?
Example
:IMPEDANCE:SUPERPOSE:FREQUENCY? ->
:IMPEDANCE:SUPERPOSE:FREQUENCY:
RANGE HZ;VALUE 100.00
5.9 IMPedance Group