Overview of the Ethernet Interface
3-3
IM 760101-11E
3
3.3
Connecting the WT to a PC
See section 13.1, “Connecting the WT1600 to a PC” in user’s manual IM760101-01E.
Ethernet Control Settings
Procedure
Select the communications interface to be used for controlling the WT.
1.
Press
MISC
to display the Misc menu.
2.
Press the
Communication
soft key to display the Comm menu.
3.
Press the
Comm Device
soft key to display the Comm Device menu.
4.
Press the
Network soft key
. Doing so selects Ethernet as the interface for
controlling the WT.
ESC
ESC
ESC
ESC
For Suffix Code -C1 (GP-IB)
For Suffix Code -C2 (Serial)
ESC
Note
Only the communications interfaces selected under Device are available. If commands are sent
using an unselected communications interface, the command will not be received.
Setting the User Name and Password
Note
When the FTP server function is specified (see section 13.1 of user’s manual IM760101-01E), the
user account and password are entered separately. We recommend that you use the same
settings as for the FTP server.
5.
Press the
User Account
soft key to display the User Account dialog box.
6.
Turn the
jog shuttle
to select User Name.
7.
Press
SELECT
to display the keyboard.
8.
Use the keyboard to enter the user name.
For instructions on keyboard operations, see section 3.12 “Entering Values and
Strings” in user’s manual IM760101-01E.
9.
Turn the
jog shuttle
to select Password. The password setting is entered twice.
10.
Press
SELECT
to display the keyboard.
11.
Use the
keyboard
to enter the password. Password is not required if the user
name is anonymous.
For instructions on keyboard operations, see section 3.12 “Entering Values and
Strings” in user’s manual IM760101-01E.
Setting the Timeout Time
12.
Turn the
jog shuttle
to select Time Out.
13.
Press
SELECT
to display the timeout time selection box.
Summary of Contents for wt1600
Page 1: ...Digital Power Meter Communication Interface IM 760101 11E 4th Edition ...
Page 131: ...7 2 IM 760101 11E 7 2 Sample Program Image ...
Page 138: ...Sample Program 7 9 IM 760101 11E 7 7 4 Output of Normal Measurement Data ...
Page 142: ...Sample Program 7 13 IM 760101 11E 7 7 5 Output of Harmonic Measurement Data ...