1-3
IM 707235-01E
Functions
1
1.3
Setting Measurement Conditions
Turning ON/OFF the Measurement Channel
Measurement is made only on the channels that have the On check box selected. If
1CH or 2CH is selected in the section “Selecting the Number of Measurement Channels”
as described later, only the specified number of channels can be selected.
Turning ON/OFF the Supply Current to the Acceleration Sensor (Bias)
When measuring acceleration, a 4-mA current is supplied to the acceleration sensors for
those channels that have the Bias check box selected. Select the
Bias
check box after
connecting the acceleration sensors.
Input Coupling
Select which component of the input signal is to be acquired.
AC: Acquire only the AC component of the input signal.
DC: Acquires all the components (DC and AC) of the input signal.
When measuring acceleration, only AC is selectable.
Measurement Unit
Set the unit for the item being measured.
When measuring acceleration
m/S2(m/s
2
)
G
When measuring voltage
V
Measurement Range/Gain
Two types of input signals can be measured: acceleration and normal voltage.
You can select from the following measurement ranges/gain.
Acceleration Measurement (Gain)
Setting
Note
×
1
Corresponds to the
±
5-V range of voltage measurement
×
2
Corresponds to the
±
2.5-V range of voltage measurement
×
5
Corresponds to the
±
1-V range of voltage measurement
×
10
Corresponds to the
±
500-mV range of voltage measurement
×
20
Corresponds to the
±
250-mV range of voltage measurement
×
50
Corresponds to the
±
100-mV range of voltage measurement
×
100
Corresponds to the
±
50-mV range of voltage measurement
Voltage Measurement (Range)
Setting
Measurable Range (Accuracy Guaranteed Measurement Range)
50 mV
±
50 mV
100 mV
±
100 mV
250 mV
±
250 mV
500 mV
±
500 mV
1 V
±
1 V
2.5 V
±
2.5 V
5 V
±
5 V
10 V
±
10 V
25 V
±
25 V
50 V
±
50 V
Acceleration Sensor Sensitivity
Enter the sensitivity of the acceleration sensor to be used. If you click