3-2
IM 707235-01E
3.2
Acceleration Sensor Connection Test
The test performs diagnosis on shorted and opened conditions of the acceleration
sensors including the cables.
CAUTION
Perform the connection test only when acceleration sensors are connected. A
current of approximately 4 mA is supplied to the acceleration sensors during the
test. Therefore, if a signal source other than an acceleration sensor is used
during the test, the signal source may be damaged.
Executing the Test
Check that the acceleration sensor is connected to the terminal. The test cannot be
performed while measurement is in progress.
1. Click
More Option
on the operation panel and select the
On
check box under
Connection Test
.
The displayed screen switches to the test mode.
Check that the check boxes for the channels being tested are selected
Select the check box
2. Check that the On check box for the channel to which the acceleration sensor being
tested is connected is selected and click
Exec
. The test is executed.
Note
The ON/OFF setting of the channel performing the connection test cannot be copied using the
copy-between-channels function.