<5. Calibration of the Instrument for Measurement Analysis>
5-4
IM 12Y18A04-01EN
1st Edition : Oct. 04, 2019-00
5.8
Auto Restart Mode
The default mode of the instrument is auto restart. In this mode, if the instrument keypad is
untouched (i.e. stays idle) for the period of auto restart idle time (default value of 30 minutes),
then it automatically starts the measurement, the mode is displayed as RUN MODE and repeats
the measurement as per the measurement frequency set in analysis parameters.
If the STOP key is pressed during this measurement, it aborts the measurement and will restart
after the idle time countdown expires.
To stop the auto restart of the instrument the mode should be changed to manual restart:
MENU → SETUP → (+) ANALYSIS PARAMETERS → MODE
5.9
Manual Restart Mode
This mode allows you to START the measurements manually.
To perform measurement on command from the MAIN MENU follow the sequence:
MENU → MEASUREMENT → ANALYZE ONCE
Pressing RUN key in this mode keeps the instrument in RUN MODE and pressing STOP key
during measurement gets back to the MANUAL RESTART mode. In this mode, if the instrument
is untouched for a period of the standby mode idle time (60 minutes default), it switches to
standby mode.
During standby operation, all fluidics devices are shut down and the analyzer goes into a standby
housekeeping mode in which the pumps are activated on an hourly basis to prevent tube
pinching.
When manual restart mode is selected, the analyzer automatically goes into standby operation
after one hour (60 minutes) of inactivity. Consult should you require a different standby time out;
for details, refer to Chapter 6.
NOTE
Even though the instrument is running continuously and displays RUN MODE, to know the actual
RESTART mode of the instrument from MAIN MENU follow the sequence:
MENU → SETUP → (+) ANALYSIS PARAMETERS → MODE
5.10 Measurement Cycle
The WE440 Phosphate Analyzer follows a sequence of steps or events for performing the
analysis cycle. They are:
1. Cell Wash
2. Sample Capture
3. Optics Testing
4. Test Reaction Temperature (when enabled)
5. Blank Measurement
6. Add reagent and stir the solution
7. Collect Data and Output Result
8. Post Wash