<3. Analyzer Operation>
3-25
IM 12Y18A03-01EN
1st Edition : Oct. 04, 2019-00
Default
Operator Action
Scrolling Marquee
Main Display
Notes
Press
to accept the setting and
return to the main setup mode
CH1 or CH2
(flashing)
SEL
SCrn
Display for two
channel analyzer
Press
to return to the channel
specific menu options in the setup
mode
Action required for
two channel analyzer
DYN
(flashing)
SEL
SCrn
SEL CH1 or SEL
CH2 on main display
of two channel
analyzer
Press
to scroll to the next menu
3.3
Shutdown and Start-Up Procedure
The following steps should be taken if a loss of sample flow is expected for more than one day.
These procedures will prevent possible build-up of caustic reagent vapors in the analyzer.
n
Shutdown
WARNING
Be sure to read and observe the following requirements.
l
Air Pump Shutdown
WARNING
Turn off the air pump prior to removing the reference electrode filling solution bottle. If the air
pump is left on, it will spatter filling solution as the bottle is removed.
1. To access air pump, press
.
2. Press /
to scroll though parameters until TEST appears in marquee window.
3. Press to access the TEST menu.
4. Press /
to scroll through the TEST menu until AIr appears in the main display.
5. Press .
6. Press /
to toggle between On and OFF settings for the air pump power. Select the
OFF setting. Changing the air pump setting to On or OFF will be indicated by a click sound.
7. Press to accept the setting and return to the TEST menu.
8. Press to return to the measure mode.
l
Analyzer Shutdown
1. Shut off sample flow prior to the analyzer inlet. Refer to Chapter 4, Flow Cell Operation.
2. Disconnect the power. Refer to Section 2.3, Wiring the Analyzer.
3. Drain the flow cell.
4. Remove the reagent bottle and store it in a well-ventilated area such as a laboratory fume
hood.
5. Carefully pull the iodide and reference electrodes out of the top of the flow cell and let them
hang by their connectors.