5-9
IM 703155-01E
Setting the Output Waveform
5
5.5
Setting the Output Shift
Procedure
To exit the menu during operation, press
ESC
located above
the soft keys.
In the procedural explanation below, the phrase “
rotary knob
& SELECT
” and “
numeric keys & ENTER
” may be used.
These phrases indicate procedures for selecting/setting items
and entering values, character strings, and units. For details
on the procedure, see section 3.7 or 3.8.
BACK
SPACE
DISPLAY
ON/OFF
KEY
LOCK
INITIALIZE
HDD
OPTION
MISC
FILE
CLOCK
MANUAL
TRIG
WAVE
SEQ
TRIG
DISP
OUT 1
OUT 2
OUT 3
OUT 4
REMOTE
UNLOCK
OFF
PRESET
SHIFT
1.
Press
WAVE
. The Wave menu appears.
2.
Press the
Next
soft key.
3.
Press any key from
OUT 1
to
OUT 4
to select the output waveform to be set.
4.
Press the
Clock Shift
soft key.
5.
Enter the output shift using
numeric keys & ENTER
. The clock shift of the
output waveform in the Wave View list is set to the specified value.
Explanation
The point at which the waveform is output when a start trigger is activated can be shifted
forward by the specified number of clocks.
Selectable range: 0 to 1000000 points (in 1 point steps)
Default value: 0 point
You can set the shift for each output terminal set (OUT1 to 4).
Converting to Shift Time
Shift time = Set point
×
(1/clock frequency)
*
For a description of the clock frequency, see section 5.3.
Note
You cannot change the output shift while waveform generation is in progress.