<6. Functions>
6-17
IM 12K01B02-01E
6.5.8
Unit Setting for TURB (turbidity)
TURB MEAS. SETTING 2/2 screen
MEAS.
screen
MAINTENANCE
screen
[MAINT.]
[MEAS.SETTING]
Refer to “6.5 Setting Screen Display for each Measuring Component”
TURB MEAS. SETTING 1/2 screen
MEAS. SETTING
screen
[TURB]
[ ]
Here describes how to set the unit for TURB.
Select one of the turbidity units from “deg.”, “mg/L”, or “NTU”.
1. Press the button for UNIT on the TURB MEAS. SETTING 2/2 screen.
The TURB UNIT screen appears.
UNIT
2. Select the unit and press [SET].
The screen returns back to the TURB MEAS. SETTING 2/2 screen.
Unit setting for turbidity is complete.
NOTE
It is common to set “deg.” when the calibration solution is PSL, “mg/L” when the calibration
solution is kaolin, and “NTU” when the calibration solution is forumajin.
6.6
Maintenance - Action
This function performs operations using the operating buttons.
Action
Description
WIPER MOTOR
Check the ON / OFF operation of the rotary ell wiper motor.
UV LAMP
Check the ON / OFF operation of the UV lamp for light.
6.6.1 Action
MEAS.
screen
MAINTENANCE
screen
[MAINT.]
[ACTION]
Refer to “6.3 Maintenance Screen Display”
ACTION screen
Here describes the execution methods for wiper motor and UV lamp.