<6. Functions>
6-5
IM 12K01B02-01E
CAUTION
Only the time within 620 minutes of current set time can be input to set the correct clock time. If
the time out of this range is input, the setting becomes invalid.
6.4.3
Output Setting
MEAS.
screen
MAINTENANCE
screen
[MAINT.]
[SETTING]
Refer to “6.3 Maintenance Screen Display”
OUTPUT SETTING
screen
SETTING 1/2
screen
[OUTPUT SETTING]
Here enables the output parameter setting.
Setting item
Description
Initial setting
ANALOG OUTPUT LEVEL Sets the analog output level.
Setting range: 0 mA to 16 mA, 4 mA to 20 mA
4 mA to 20 mA
MAINT.OUTPUT
Sets the measurement output during calibration/maintenance
to output by ont of the settings below.
Setting range: HOLD, DIRECT, PRESET (0 mA to 21 mA)
HOLD
WARM-UP OUTPUT
Sets the measurement output during warm-up to output by
one of the settings below.
Setting range: HOLD, DIRECT, PRESET (0 mA to 21 mA)
PRESET
4.0 mA
ALARM OUTPUT
Sets the measruement output during an alarm to output by
one of the settings below.
Setting range: HOLD, DIRECT, PRESET (0 mA to 21 mA)
HOLD
1. Press [OUTPUT SETTING] on the SETTING 1/2 screen.
The OUTPUT SETTING screen appears.
2. Press the button for ANALOG OUTPUT LEVEL.
The ANALOG OUTPUT LEVEL screen appears.
3. Select the setting item from [0 - 16] or [4 - 20] and press [SET].
The screen returns back to the OUTPUT SETTING screen.
Setting for the analog output level is complete.
l
Output setting during operation
Here describes the following three output setting methods.
• MAINT. OUTPUT
• WARM-UP OUTPUT
• ALARM OUTPUT
MEAS.
screen
MAINTENANCE
screen
[MAINT.]
[SETTING]
Refer to “6.3 Maintenance Screen Display”
OUTPUT SETTING
screen
SETTING 1/2
screen
[OUTPUT SETTING]
Summary of Contents for UV700G
Page 5: ...Blank Page ...
Page 15: ...Blank Page ...
Page 35: ...Blank Page ...
Page 39: ...Blank Page ...
Page 115: ...Blank Page ...
Page 117: ...Blank Page ...
Page 131: ...Blank Page ...