2. Specification
2.1
Standard Specification
Measurement :
Organic pollutants in wastewater.
Measurement method:
Dual optical path, dual wavelength, rotary cell length modulation method
Measurement wavelength:
Ultraviolet (UV):
253.7 nm
Visible light (VIS):
546.1 nm
Light source:
Low pressure mercury-vapor lamp
Detector:
Silicon photo cell, flow-through type
Measuring range (corrected for 10 mm cell)
UV/VIS absorbance 0-0.1 to 0-5.0 Abs
Display:
Graphic LCD with touch screen,
320
240 dot, with backlight
Indication:
UV absorbance, VIS absorbance, UV- VIS absorbance, COD
concentration, turbidity
Resolution:
0.0001 Abs (factory default) or 0.001 Abs
COD conversion:
Conversion equation
Y = a
bX
where
Y
= COD concentration
X
= absorbance (UV-VIS)
Turbidity conversion: Conversion equation
Y = a
bX
where
Y
= turbidity,
X
= absorbance(VIS)
Analog output:
Number of outputs: 3
Functions: 3 functions selectable:
UV absorbance / VIS absorbance/ UV-VIS absorbance / COD
concentration / turbidity
Specifications:
4-20 mA DC or 0-16 mA DC, isolated
(non-isolated between channels)
Contact output:
Number of outputs: 6
Functions:
2 functions fixed:
Power off, maintenance status
4 functions selectable:
Total alarm / COD high alarm / turbidity high alarm / light souse
malfunction / no sample / cleaning motor error / analyzer error
Descriptions:
Power off:
Activated when power off occurs
Maintenance status:
Activated when instrument is in maintenance or calibration mode or
when MAINTENANCE switch is turned on
Total alarm:
Cleaning motor error, light source malfunction and / or analyzer error