<9. Troubleshooting>
9-2
IM 12K01B02-01E
9.3
Alarm Occurrence Condition
Item
Alarm
Alarm triggered
Alarm cleared
Alarm release
*3
Common
WARM-UP
For 60 minutes after power ON
After 60 minutes passed since power ON
CELL MOTOR ERR.
When a rotation of the motor takes 5
seconds or more.
When a rotation of motor takes less than
5 seconds.
SENSOR COMM.
ERR.
*1
When detector communication
is abnormal.
When detector communication
resumes.
SAMPLE LACK
When contact input of no sample is
ON for 5 seconds or more.
When contact input of no sample is OFF
for 5 seconds or more.
PIO COMM. ERR.
*2
When communication of the input/
output substrate is abnormal
When communication of the input/output
substrate resumes.
POWER ON/OFF
History is saved at start up.
BATTERY ERR.
When the backup battery voltage for
clock is less than 2.4 V.
When the backup battery voltage for
clock is 2.4 V or over.
LAMP ERR.
When the light intensity of the light
source is decreased.
When the light intensity of the light
source is restored.
UV
ZERO CAL.
When zero calibration value is out
of the calibration range.
When zero calibration value is within the
calibration range.
SPAN CAL.
When span calibration value is out
of the calibration range.
When span calibration value is within the
calibration range.
VIS
ZERO CAL.
When zero calibration value is out
of the calibration range.
When zero calibration value is within the
calibration range.
SPAN CAL.
When span calibration value is out
or the calibration range
When span calibration value is within the
calibration range
COD
HIGH
The status "measurement value >
MAX ALARM VALUE" continues
longer than the specified alarm time
The status "measurement value < MAX
ALARM VALUE" continues longer than
the specified alarm time
LOW
The status "measurement value <
MIN ALARM VALUE" continues
longer than the specified alarm time
The status "measurement value > MIN
ALARM VALUE" continues longer than
the specified alarm time
TURB
(turbidity)
HIGH
The status "measurement value >
MAX ALARM VALUE" continues
longer than the specified alarm time
The status "measurement value < MAX
ALARM VALUE" continues while
exceeding the specified alarm time
LOW
The status "measurement value <
MIN ALARM VALUE" continues
longer than the specified alarm time
The status "measurement value > MIN
ALARM VALUE" continues longer than
the specified alarm time
SPAN CAL.
When span calibration value is out
of the calibration range.
When span calibration value is within the
calibration range.
*1 SENSOR COMM.ERR.:
Communication error between detector substrate and
converter substrate.
*2 PIO COMM.ERR.:
Communication error between converter CPU substrate
and input / output substrate.
*3 Alarm release:
Forced stop by pressing [DELETE] on the ALARM screen.
: Forced stop is possible.
: Though forced stop is possible, the alarm recurs because
the cause has not been resolved.