<5. FLXA402T sensor menu>
5-5
IM 12E04B40-02EN
1st Edition : Mar. 08, 2023-00
5.4
Reset op. time
To reset wellness data of each sensor
Sensor menu > Reset op. time > Sensor wellness , tap “Reset data” > Reset data
Figure 5.6
You can reset each data from the dialog of Sensor wellness. If you reset, the operating days are
reset to “0” but the replacement date is updated.
When you have a password, the password verification appears. Upon successful confirmation,
Reset confirmation screen appears.
For further information on the password setting, see FLXA402T Operation of Converter IM
12A01G01-03EN 5.4 Password.
5.5 Setting
Sensor menu > Setting .
Figure 5.7
When you have a password, the password verification appears. Upon successful confirmation,
Sensor Setting screen appears.
For further information on the password setting, see FLXA402T Operation of Converter IM
12A01G01-03EN 5.4 Password.
If you go to Sensor setting, the mA output becomes HOLD when automatic hold function is
enabled, and contact output keeps the current status.
n
Sensor setting
Set sensors. See “4. SETTING PARAMETERS”
n
Initialization
Initialize sensor parameters.
If you tap “Execute”, the loading starts. When the loading ends, you will return to Sensor menu.
Confirm the items before tapping Execute. If you initialize the calibration value of Zero, Sens.
factor, linearize table etc., they return to the factory default. Some factory default are different