< 4. SETTING PARAMETERS >
4-8
IM 12E04B40-02EN
1st Edition : Mar. 08, 2023-00
4.5.4
Auto sequence setting
Configure the following parameters to setup the schedule for Automatic cleaning and Automatic
zero calibration.
• Cleaning interval
• Repeat count of wash
• Calibration frequency
The auto-sequence starts at the setup hour of the start time at the washing interval cycle.
See how many times the automatic cleaning repeats in a sequence by setting “Repeat count of
wash”. When Auto calibration function” is ON, Auto sequence starts at an interval determined
calibration frequency. When automatic cleaning ends, automatic zero calibration operates one
time. When Calibration frequency is set to “0”, no Auto calibration operates.
The next figure shows the example when. Repeat count of wash is 2. Calibration frequency is 3.
Start date and time
Next sequence start
Next sequence start
Time schedule
Washing interval
Number of washes
Washing interval
Number of washes
Washing interval
Number of washes
Wash Wash
Wash Wash
Wash Wash
Cal
Wash
Wash
Automatic calibration takes place one time after the auto sequence repeats as many calibration frequencies.
Other parameters are for setting details about one count of automatic cleaning/automatic zero
calibration. For further information on the auto sequence, read “Appendix Automatic cleaning/
Automatic zero calibration Sequence”.
CAUTION
The settings changed in the wash interval will be reflected in the second “start date and time” that
follows.