<3. PREPARATION for OPERATION>
3-5
IM 12E04B40-02EN
1st Edition : Mar. 08, 2023-00
3.11 Zero calibration & slope calibration
After performing the running-in and the display becomes stable, perform zero calibration using
zero water and span calibration using a calibration plate.
Refer to “7. CALIBRATION” for calibration procedure.
It is recommended to perform calibration approximately once a month. Calibration must be
performed when the product has been out of operation for a long time or after inspection or
maintenance.
3.12 Supplying Sample and Adjustment flow rate
(1) Turn Off converter maintenance mode.
(2) Set the manual valve to the measurement state (see
(3) Supply the specified flow rate (2-10 L/min) of the sample.
(4) Measure the displacement of the detector drain with a graduated cylinder, etc., and confirm
that the displacement is within 1.5 L ± 0.1 L per minute. If the flow rate is out of range,
perform horizontal adjustment, or flow rate adjustment) of the detector referring to
“3.6 Leveling detector (flow rate adjustment)”
.
(Note 2) For the specification without a sampling system (sampling system code:-NN,-A5), make sure that the flow rate of the sample
flowing into the detector is 1.5 to 2 L/min .
CAUTION
Precautions for products when option /PH□2, /PH□7, or /FC is specified.
The pH sensor of the pH meter and the ceramic beads of the Free Chlorine Meter detector
are shipped in separate packages. Refer to the user’s manual of each instrument and install it
correctly as a preparation for operation.
Failure to do so may cause measurement failure or equipment failure.
The operation of the instrument begins with supplying the required process fluid to the respective
device. Refer to the piping diagram in “2.1 Piping”
and adjust to the following flow rate by
operating the appropriate valve.
(1) 3 to 11 L/min for pH meter Flow-through Type Holder
(2) Apply 0.1 to 2.5 L/min to the non-reagent type free chlorine detector.
For the operation of each device, refer to the respective instruction manuals.
3.13 Switch to measurement mode
After operation preparation is completed, turn Off the maintenance mode and set it to
measurement mode. Check that no error occurs.