<2. INSTALLATION, PIPING, AND WIRING>
2-8
IM 12B01B30-02EN
1st Edition : Mar. 08, 2023-00
Power
Detector
Power
L
N
G
Power
RS485(-R)
Ethernet(-E)
SENCOM
RELAY
CN1
Converter
Measure
Wash
Zero
Drain
O
S
C
SV1 (-A3)
SV2 (-A2, -A3)
SV3 (-A3)
SV4 (-A2, -A3)
+
–
+
–
+
–
+
–
+
–
(61)
(62)
mA1
(65)
(66)
mA2
(81)
(82)
mA3
(85)
(86)
mA4
(21)
(22)
(24)
SHLD (63)
(87)
(88)
SHLD (89)
A+ (91)
B
−
(92)
GND (93)
SHLD (94)
Termination (95)
Sensor1
Sensor2
(32) NC
(31) C
(33) NO
S1
(42) NC
(41) C
(43) NO
S2
(52) NC
(51) C
(53) NO
S3
(72) NO
(71) C
(73) NC
S4
L (101)
N (102)
G (001)
(11)
(12)
(21)
(22)
(31)
(32)
(41)
(42)
(13)
L (101)
N (102)
G (001)
L (201)
N (202)
G (002)
L (301)
N (302)
G (003)
RELAY BOX
FLXA402T
Liquid Analyzer for Turbidity and Chlorine
Power
Supply
Power
Supply
Power
Supply
TB830D
Surface Scattered Light Turbidity Detector
Analog output (-N2, -N4)
Analog output (-N2, -N4)
Analog output (-N4)
Analog output (-N4)
Contact input
(-N2, -N4)
Analog input (-N4)
Contact output
(-WR)
L (101)
N (102)
G (001)
L (201)
N (202)
G (003)
L (101)
N (102)
G (001)
Relay box
B
L (101)
N (102)
G (001)
L (201)
N (202)
G (003)
L (101)
N (102)
G (001)
CP
AR
Relay box A
Power supply cable (E)
Power supply cable (E)
Relay Box power supply cable (D)
• If a Terminal box is not included, wire power supply cable (E) to the Relay box.
• If Terminal box A or B is included, wire the power supply cable (E) to the Terminal box.
• Wire the Relay box power cable (D) between the Terminal box and the Relay box
Figure 2.8
Wiring TB830D-□-□□A2, -A3
SENCOM cable for
Turbidity/Chlorine (A)
FLXA402T
power supply cable (B)
TB830D power supply cable (C)
or power supply cable (E)
Clamp
Never let the power cables of TB830D touch the
SENCOM cable for turbidity/chlorine (A), resulting in
failure to comply with the safety requirements.
• Open detector cover to wire.
• Wire SENCOM cable for turbidity/residual salt through the clamps in the figure. Never let the power cables of TB830D touch
the SENCOM cable for turbidity/chlorine (A), resulting in failure to comply with the safety requirements.
• A cable gland is attached to the wiring port. Securely tighten the wires so that they meet the IP65 or higher.
• After wiring is complete, securely tighten the four screws on the detector cover. The tightening torque of the screw is 1.5 N·m.
Figure 2.9
TB830D WIRING
WARNING
Wire power supply cable (TB830D power supply cable (C) or power supply cable (E)) in Figure
so that it does not come into contact with the SENCOM cable. Otherwise, it results in failure
to comply with the safety requirements.