<8. MAINTENANCE>
8-6
IM 12B01B30-02EN 2nd Edition : June 09 , 2021-00
8.6
LED Replacement
CAUTION
•
Turn off the power of TB820D before replacing the LED.
• LED may be damaged by static electricity. When replacing the LED be sure to take
measures against static electricity, such as using a grounded wristband.
Although the LED has a normal service life of more than three years, it is recommended that the
LED be replaced at an interval of three years in terms of preventive maintenance. Some LEDs
may fail within three years.
The following describes how to replace the LED.
(1) Stop power supply to TB820D.
(2) Remove the front cover of the detection unit by loosening the six (6) setscrews.
(3) Loosen the 1 setscrew of LED bracket.
(4) Rotate upward the LED bracket and remove.
(5) Detach the cable connector from the LED bracket.
(
See Figure 8.6)
(6) Replace the LED bracket with a new one.
(7) Attach the new LED bracket to the cable connector.
(
See Figure 8.7)
(8) Rotate downward the LED bracket to set.
(9) Fasten the 1 setscrew to fix the LED bracket.
(10) Fasten the six (6) setscrews on the front cover of the detection unit and attach the cover.
CAUTION
After turning on the power of the instrument, warm it up for at least 30 minutes before starting the
measurement.
Arrange the LED as below.
Table 8.4
LED
Part No.
Name
Description
K8003DN
LED (red light)
for -NT, 660 nm Light source
K8003DP
LED (infrared light)
for -FN, 860nm Light source