< 7. Calibration >
7-11
IM 12E01B30-02EN 2nd Edition : June 09 , 2021-00
7.10 Basic (reference) sensitivity calibration
Sensitivity factor(S0) is updated The slope is 100 %. Be sure to perform this calibration after
implementing the Zero calibration.
No error check is conducted at the calibration, therefore, perform calibration properly. For
information on preparation of calibration solution, see “
[How to perform basic (basic (reference)) sensitivity calibration]
1
Turn on the maintenance mode.
2
Press on Main screen to enter the calibration menu.
3
Select [basic sensitivity calibration] in the menu list of Calibration others.
4
Remove the top cover from the top of the detector. Loosen the four screws fixing the clear
cover and remove the clear cover.
5
Use a funnel and pour standard solution into the Flow cell to a little higher than the level
of the measurement window. When the standard solution is supplied to the level, open
the drain valve and drain the solution which has been poured. Pour the solution again
slowly. Do not allow bubbles enter the cell.
When the Flow cell is filled with the standard solution to the level which is a little higher
than that of the measurement window, put the top cover in place.
6
Confirm the reading stable. Press [check stability].
7
When a message appears to notice the stabilization is confirmed, enter a calibration
value.
Press [Next]. As the calibration value, enter a manually acquired value in the step 2
above.
8
Check the calibration result. Press [Accept] to update a basic (basic (reference))
Sensitivity factor(S0) of calibration parameter. Slope is 100%.
9
To exit the calibration, turn off the maintenance mode.
7.11 Calibration Error
When executing calibration items except for Basic sensitivity calibration, if a calibration result
is abnormal, the calibration error is displayed, and the calibration result is discarded. The
Acceptable range as follows:
Zero:
Acceptable range is wide and hardly error occurs.
Slope: 25~200%
Zero shift value:
-10 ~ 10 NTU
Correction factor:
0.25~4 times