<2. PIPING AND WIRING>
2-2
IM 12E01A06-02E
2.1.2 System without Head Tank and Zero Turbidity Filter
This is a simple system where a water sample is taken from the process and directly introduced
into the detector. This system configuration can be used when a water sample contains a negligible
amount of air bubbles or when the turbidity of a water sample is high and the effect of air bubbles is
nonsignificant.
CAUTION
Sample water conditions and ambient temperature must meet the specifications: maximum pressure:
500 kPa; sample temperature: 0 to 50°C; ambient temperature: -5 to 50°C.
h
F2-2E.ai
Sample Water
Inlet
Drain Port
Sample Water
Outlet
TB750G
Detector
Drain Valve
Sample
Water
Tap Water
Sample Water
Valve
Zero Turbidity
Filter
Sample Water
Supply Valve
Zero Water Supply Valve
Tap Water
Valve
Drain
Drain
Sample Water
Zero Water
Flow Control Valve
Sample Water
Effluent
Zero Water
Drain Valve
Other Turbidity Detectors
Process Piping
Figure 2.2
Piping Diagram
• The flow control should not be done on the inlet side of the detector.
• To prevent corrosion inside the zero turbidity filter, allow water to flow through the filter and flow
out from the zero water drain valve continuously at a flow rate of approximately 10 ml/min.