<3. OPERATION>
3-2
IM 12E01A06-02E
3.1.3 Calibration
(1) When performing a calibration using 0.2 or 1.0 µm filtered water as zero reference, refer to
section 3.1.4.
(2) When performing a 2-point calibration using standard solutions, refer to section 3.1.5. Use this
method to perform a calibration complying with EPA 180.1.
Note: The instrument has been pre-calibrated using 0.2 µm filtered water as zero reference at the
factory before shipment.
Calibration Flow Chart
F3-3E.ai
YES
Check block
Standard Solution
NO
* Calibration complying with EPA Method 180.1.
Is filtered water used
as zero reference?
Is span calibration performed
using check block or
standard solution?
Perform span/zero calibration
using filtered water
Perform 2-point calibration
using standard solutions*
Perform zero calibration
Perform point calibration
first using zero solution
Perform span calibration
using check block.
Refer to Section 3.1.4.
Perform span calibration
using standard solution.
Refer to Section 3.1.4.
Perform second point
calibration using span solution.
Refer to Section 3.1.5.
Start calibration
End calibration