5-1
IM 704610-01E
Displaying Measured Results
5
Chapter 5
Displaying Measured Results
5.1
Meter Display
<<For a functional description, see section 2.5>>
Procedure
MEDIA
FUNCTION
POLARITY
ARMING
INHIBIT
LOCAL
RECALL
ENTER
TRIG
IMPEDANCE EQUALIZER
SEC RANGE
KEY
LOCK
MEASURE
DISP OFF
JUDGE
GO
NO-GO
10%
20%
S
SEC RANGE
SCALE
L.MARK
R.MARK
GATE TIME
DELAY
BLOCK
REMOTE
STORE
UTILITY
JDG LEVEL
LEVEL
AUTO
COUPLE
EQ BOOST
MAN
CONV
50
Ω
1M
Ω
PW
E2T
D to C
BD x1
ns ms
s
%
V
%/
S
SHIFT
SHIFT
Display 1
Display 2
The phrase, “rotary knob & <>”
is used in the following
explanation to instruct the user
to enter numerical setting values
using the knob and arrow keys.
For details on this procedure,
see section 3.7.
Displaying the Time Range
If you press the SCALE key and select s, the units of jitter displayed on the meter are set to units
of time (ns or
µ
s). In this case, the specified time range (SEC RANGE) is displayed here.
Displaying the Jitter Ratio and Selecting the Scale
Press SCALE to select 10% or 20%. The needle moves according to the
selected scale. The jitter ratio can be read on the upper scale of the meter (%).
Displaying the Jitter and Selecting the Time Range
1.
Press SCALE to select s.
2.
Press SHIFT+SCALE (SEC RANGE). The currently set time range is displayed
in the SEC RANGE area of Display 2.
3.
Turn the rotary knob to select the time range. The selected time range is
applied to the lower scale (s) on the meter, and the needle moves according to
the new range. The jitter can be read on the lower scale of the meter (s).
Explanation
The TA220 continuously takes measurements when the power is turned ON. The meter
needle indicates the jitter ratio or jitter of the measurement function that was selected in
section 4.1.
Jitter Ratio Scale
You can select the 10% or 20% scale for the jitter ratio on the upper scale of the meter.
• 10%
The scale line is written every 0.2%. The needle is capable of indicating a jitter ratio of
up to 11%.
• 20%
The scale line is written every 0.5%. The needle is capable of indicating a jitter ratio of
up to 22%.
Jitter Scale
The lower scale on the meter (s) is used as the jitter scale.
Time Range
When the scale units for the meter are set to s, you can select a time range from the
following.
0.5 n, 1.0 n, 5.0 n, 10 n, 50 n, 0.1
µ
, 0.5
µ
, 1.0
µ
, 5.0
µ