13-6
IM 704610-01E
13.7 General Specifications
Item
Specifications
Standard operating conditions
Ambient temperature:
23
±
5
°
C
Ambient humidity
50
±
10%RH
Error in supply voltage
Within 1% of rating
and frequency
Warm-up time
Approx. 30 minutes.
Storage Environment
Temperature
–20
°
C to 60
°
C
Humidity
20% to 80% RH (no condensation)
Operating Environment
Temperature
5
°
C to 40
°
C
Humidity
20% to 80%RH (no condensation)
Rated supply voltage
100 to 120 VAC, 200 to 240 VAC
Permitted supply voltage
90 V to 132 VAC, 180 V to 264 VAC
range
Rated power supply frequency 50/60 Hz
Allowable power supply
48 to 63 Hz
frequency variation
Maximum power consumption 150 VA
Insulation resistance (between power supply and case)
10 M
Ω
or more (500 VDC)
Withstanding voltage (between power supply and case)
2500 VAC at 50/60 Hz for 5 s
Signal ground
The ground of all input and output connectors are connected to the case ground. In addition, all
input terminals are protected using a diode.
External dimensions
Approx. 213 (W)
×
132 (H)
×
350 (D) mm excluding projections.
Weight
Approx. 5 kg (main unit only)
Cooling method
Forced air
Installation position
Horizontal (stacking prohibited)
Battery backup
Setup information is backed up with the internal lithium battery
Keylock
Able to set key lock.
Recommended calibration
1 year
interval
Safety standard
Complying standard EN61010-1
• Overvoltage category (Installation category) II*1
• Pollution degree 2*2
• Measurement category II*3
Emission
Complying standard EN61326 Class A, C-Tick AS/NZS CISPR11
This product is a Class A (for commercial environment) product.
Operation of this product in a residential area may cause radio interference
in which case the user is required to correct the interference.
Cable requirement
• BNC Connector
Use a BNC cable*4.
• GP-IB interface connector
Use a GP-IB shielded cable*4.
• Ethernet interface connector
Use an Ethernet interface cable*5.
Tests
1. Main terminal disturbance voltage (EN61326) Class A
2. Electromagnetic radiation disturbance (EN61326) Class A
3. Power supply harmonics restriction (EN61000-3-2)
4. Voltage fluctuation and flicker (EN61000-3-3)