12-7
IM 704610-01E
Troubleshooting and
Maintenance
12
Meter Test
3.
Press ARMING . METER is displayed in Display 2. Confirm that the needle
points almost to the center of the scale.
4.
Turn the rotary knob to the left and right. Confirm that the values displayed in
Display 1 increase or decrease, and that the needle moves to the left or right
accordingly.
5.
Press <. TESTMODE is once again displayed in Display 2.
Board Test
3.
Press INHIBIT. BOARD is displayed in Display 2, the sequence of tests is
displayed in Display 1 as they are carried out (Cpu -> MEAS -> inPut -> EqLZ ->
EtHEr), and then PASS or FAIL is indicated.
If PASS is displayed, the board is operating normally. Go to step 5.
If FAIL is displayed, this indicates a problem with the instrument’s port. Go to
step 4.
Confirming the Test Results of Each Board
4.
Turn the rotary knob to display CPU b., MEAS b., INPUT b., EQLZ b., or
ETHER b. in Display 2. PASS or a hexadecimal value is displayed in Display 1.
If PASS is displayed, the board is operating normally.
If a hexadecimal value is displayed, a problem has occurred. Please note the
hexadecimal value that is displayed.
The characters displayed in Display 2 correspond to the following boards.
CPU b.: CPU board
MEAS b.: Measure board
INPUT b.: Input board
EQLZ b.: Equalizer board
ETHER b.: Ethernet board
To perform all board tests again, press INHIBIT.
5.
Press <. TESTMODE is once again displayed in Display 2.
Note
When you wish to perform measurement again after the self test, reboot the TA220.
12.4 Self Diagnostics (Self Test)