9
TI 01W06A51-50EN
4.3
Installing the Gateway
Attach the gateway to a pipe in the installation location. If there are no appropriate pipes, prepare
one in advance. A flat blade screwdriver is required to carry out this work.
Check that the following items have been prepared.
•
Flat blade screwdriver
•
Stainless hose band (x 2)
•
Gateway with the bracket mounted
CAUTION
• To mount the gateway, use a pipe with a diameter of 50 to 125 mm.
• The hose bands provided with the gateway can be used for a pipe with a diameter of 78 to
101 mm. When using a pipe that is any larger or smaller, prepare appropriate hose bands
separately.
• When installing the gateway in an elevated location, watch your step.
Follow the procedure below to carry out the work.
(1) Prepare the two stainless hose bands that came with the gateway, and pass each of them
through the gateway bracket.
(2) Loop the hose band, and pass the mounting pipe through the loop. Tighten the screw of the
hose band using a flat blade screwdriver. As the screw of the hose band is tightened, the
hose band is secured.
(3) Be sure to use two hose bands to secure two sections of the mounting pipe.
Diameter: 50 to 125 mm
Stainless hose band
(accessory)
Mar.26,2019-00