Setup Operations for Changing/Adding Functions (Basic Setting Mode)
4
4.10 Enabling the Moving Average Function (Dot
Model)
The moving average function can be enabled/disabled on the dot model. Basic Setting
mode cannot be entered when recording is in progress.
Procedure
1.
Hold down the
MENU
key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to display the
Basic Setting mode screen.
3.
Carry out the procedure shown in the figure below.
Press the
or
key to select the value.
If you press the
key, the operation is cancelled, and the display returns to a
higher level menu.
The settings are activated.
Select whether to use moving average
function.
Select “M_AVE.”
(M_AVE)
(NOT)
(OK)
4.
If you are done, press the
key.
5.
To return to the Operation mode,
1. Press the
key, use the
or
key to select
, and press the
key.
2. Press the
or
key to select
or
, and press the
key.
Explanation
Moving Average
Enable (
): Enables the setting of the number of samples of the moving average
in Setting mode.
Disable (
): The “M_AVE” item does not appear in Setting mode.
<Related Topics>
Setting the moving average: section 3.3