6.17 Setting the Wave Window Trigger (ENHANCED)
<For a description of this function, refer to page 2-23.>
Procedure
1.
Press
MENU
.
2.
Press the
TRIGGER
soft key.
3.
Press the
Trigger Cond. Setup
soft key.
4.
Press the
Select
soft key to select Enhanced.
Setting the Trigger Type
5.
Press the
Type
soft key. The trigger type selection menu appears.
6.
Use
jog SELECT
to select Wave Window.
Setting the Trigger Conditions
7.
Press the
Set Pattern
soft key. The Wave Window trigger setup menu appears.
8.
Use
jog SELECT
to select the pattern of each channel (CH1 to CH16)
from ON and OFF.
Note
•
The wave window trigger does not apply to the 701265 (TEMP/HPV) and the 701280
(FREQ).
•
Wave window trigger is not applicable when measuring temperature on the 701261
(UNIVERSAL) or 701262 (UNIVERSAL (AAF)).
•
Wave window trigger cannot be used in Recorder mode.
Setting the Cycle Frequency, Reference Cycle, Synchronization Channel,
Level, and Hysteresis
8.
Use
jog SELECT
to set Cycle Frequency. Pressing
RESET
resets the
value to 50 Hz.
10.
Likewise, set the Reference Cycle to 1, 2, or 4.
11.
Likewise, set Sync. Ch to Auto or CH1 to CH16.
12.
If you select a setting other than Auto in step 11, set the Level of the
synchronization channel.
13.
Likewise, set Hys to
,
, or
.