6-12
IM 701240-01E
6.7
Activating Triggers on the Power Signal
(SIMPLE)
<For a description of this function, refer to page 2-17.>
Procedure
1.
Press
MENU
.
2.
Press the
TRIGGER
soft key.
3.
Press the
Trigger Cond. Setup
soft key.
4.
Press the
Select
soft key to select Simple.
Setting the Trigger Source
5.
Press the
Source
soft key. The trigger source selection menu appears.
6.
Use
jog SELECT
to select Line.
Setting the Hold Off
7.
Set the hold off time according to the procedures given in section 6.4.
Explanation
A triggers can be activated on the rising edge of the power signal that is being supplied
to the SL1400. Waveforms synchronized to the commercial power supply frequency (50
Hz or 60 Hz) can be observed.
Trigger Source
Select Line. The power signal that is being supplied to the SL1400 becomes the trigger
source.
Hold Off
See section 6.4.