6-6
IM 701240-01E
6.4
Setting the Hold Off Time
<For a description of this function, refer to page 2-26.>
Procedure
1.
Press
MENU
.
2.
Press the
TRIGGER
soft key.
3.
Press the
Trigger Cond. Setup
soft key.
4.
Press the
Hold Off
soft key.
5.
Turn the
jog shuttle
to set the hold off time. Pressing
RESET
resets the value
to 0.0
µ
s.
Note
The hold off time setting applies to both simple and enhanced triggers.
Explanation
This function prevents a trigger from being activated for a specified time, even if the
trigger conditions are met during this time. This is effective when you want to activate the
trigger in sync with a periodic signal as shown in the figure below.
Input signal
Repetition period:T
Trigger level
Trigger source
signal
t
Trigger signal limited by hold off time t (when the trigger slope is set to rising edge)