![YOKOGAWA ScopeCorder DL350 User Manual Download Page 81](http://html.mh-extra.com/html/yokogawa/scopecorder-dl350/scopecorder-dl350_user-manual_911346081.webp)
4-59
IM DL350-01EN
Trigger Display (Error Trigger)
If an error is detected, the pulse waveform is displayed for two sampling periods of the SENT module.
Error Count (Error Count)
An error is counted when any of the error types is detected. This shows the total number of errors.
Maximum count: 65535
Error Count Reset (Error Count Reset on Start)
Set whether to reset the error count to zero when waveform acquisition is started with the START/STOP key on
the front panel.
ON: The count is reset.
OFF: The count is not reset.
Manual Error Count Reset (Error Count Reset)
You can manually reset the error count to zero.
Input Settings (Input Setup)
Set the probe attenuation and the timeout value for SENT signal input. The threshold level is fixed.
Probe Attenuation (Probe)
Select the attenuation of the probes connected to SENT ports.
1:1, 10:1
Threshold Level (Threshold H, Threshold L)
The threshold level for determining whether the SENT signal is high or low level is fixed.
Threshold H: 3.5 V
Threshold L: 1.5 V
* The SENT module voltage range is fixed at ±20 V.
Timeout Value (Time Out)
Set the timeout value for SYNC/CAL pulse detection.
If a SYNC/CAL pulse is not detected within the specified time, all Fast CH and Slow CH data are set to zero.
Selectable range: 0.1 ms to 2000.0 ms
Resolution: 0.1 ms