![YOKOGAWA ScopeCorder DL350 User Manual Download Page 210](http://html.mh-extra.com/html/yokogawa/scopecorder-dl350/scopecorder-dl350_user-manual_911346210.webp)
19-4
IM DL350-01EN
Event Search (Event)
You can search for an event number that was assigned during measurement.
Event Number (Select Number)
Select the event number that you want to search for.
When searching manual events, the selectable range is 1 to 100 (maximum number of input events).
* When the trigger mode is auto, event search is performed only on the event numbers remaining on the
display screen.
Executing a Search (Execute)
The instrument displays the waveforms of the area around the selected event number expanded in the zoom
window.
Logic Pattern Search (Logic Pattern)
The instrument searches for the specified logic pattern. This setting is valid only for the logic input terminals (CH5
and CH6) of this instrument and logic modules.
Search Conditions (Setup)
Set the following search conditions: the waveforms to search, bit settings, and count.
Source Waveform (Trace)
Select the waveforms to search from the options listed below.
CH1 to CH6
*
* You can select only the following channels.
• Channels of logic modules
• Sub channels of CAN bus monitor or CAN & LIN bus monitor modules whose data types are set to Logic.
• S&C and Error trigger sub channels of SENT monitor modules
Bit Settings (Bit Setting)
To set the logic signal pattern that you want to detect, set the signal states for Bit1 to Bit8. The instrument
searches based on the AND of each bit.
• H: High level
• L: Low level
• X: The state of the bit is not used as a condition.
Count (Count)
Set the number of times the specified pattern must repeat. You can select a number from 1 to 1000000.
Search Range (Start Point/End Point), Search Number, Executing a Search (Execute)
19 Searching Waveforms