18-1
IM DL350-01EN
18 Waveform Display Position and Zoom (Recorder
mode only)
You can magnify displayed waveforms along the time axis. The zoomed waveform of one locations can be
displayed. You can also specify which channel you want to zoom in on. You cannot zoom if the number of
displayed points on the screen is less than or equal to 100.
Zoom window
Zoom box
Example of the Zoom Display
Main window
Vertical Zoom
You can magnify displayed waveforms along the vertical axis by using the Channel menu.
Turning the Zoom Window Display (Display) On and Off
Set whether to display the zoom window. When a zoom window is displayed, a zoom box appears in the main
window. It indicates what part of Main window is being zoomed in on.
• ON: The zoom window is displayed.
• OFF: The zoom window is not displayed.
Time Range of the Zoom Window
Set the zoom factor in time range. The zoom-window time scale changes automatically based on the specified
zoom factor.
Selectable Range
Two times the time scale (record time) of the main window to the point where the number of data points in the
zoom window reaches 10 points per div.
• When an external clock signal is being used as the time base, you can select one of the following
magnifications. You can select any magnification up to the point where the number of data points in the
zoom window reaches 10 points per div.
Up to 1000000 in 1-2.5-5 steps (×2, ×2.5, ×5, ×10, ×25, ×50, ×100, ×250, ×500, ×1000, ×2500, and so on)
• The upper limit of the zoom factor during SD recording is the maximum zoom factor that can be displayed
during SD recording.