Setting the Delay (Delay Setup)
The time difference between traces or the time difference from the trigger point to a rising or falling edge is called
the delay between channels.
: (rising)
: N1 (an integer
between 1 and 9)
: (falling)
: N2 (an integer
between 1 and 9)
Reference waveform
(Reference)
Source waveform
(Measure)
Count 1
Count N1
Count N2
Count 1
Delay between channels
• Polarity
• Edge Count
Reference waveform settings
Source waveform settings
Measurement range
• Polarity
• Edge Count
Mesial line
• Mode
Select a delay measurement mode.
• OFF: Delay measurement is not performed.
• Time: The delay between channels is displayed as a time.
• Degree: The delay between channels is displayed as an angle.
• Polarity
Select the slope of the edge you want to detect.
•
: Rising
•
: Falling
• Edge Count
Sets which edge counted from the start point (T Range1) of the measurement time period to use as a
detected point (measured point).
Selectable range: 1 to 9
• Reference
Select whether to use a trace or trigger as the reference for the reference waveform.
• Trace: A trace is used.
• Trigger: A trigger is used.
• Reference Waveform (Reference Trace)
When Reference is set to Trace, set the reference waveform.
• Trace: Select a reference waveform. CH1 to CH6
1
, 16chVOLT
2
, 16chTEMP/VOLT
2
, CAN
3
, LIN
3
, SENT
3
,
Math1 to Math2
1 You can select the channel of an installed module. On a 4-CH module, select sub channel 1 or 2.
2 When a 16-CH Voltage Input Module or 16-CH Temperature/Voltage Input Module is installed. After you
select 16chVOLT or 16chTEMP/VOLT, select a sub channel.
3 On a model with the /VE option when a CAN bus monitor, CAN & LIN bus, or SENT monitor module is
installed. After you select CAN, LIN, or SENT, select a sub channel. This cannot be selected on a CAN
bus monitor or CAN & LIN bus monitor module if the data type (Value Type) is set to Logic. On a SENT
monitor module, S&C and Error Trigger sub channels cannot be selected.
• Polarity: Select the slope of the edge you want to detect ( : rising, : falling).
• Edge Count: Sets which edge counted from the start point (T Range1) of the measurement time period to
use as a detected point (reference point). Selectable range: 1 to 9
10 Automated Measurement of Waveform Parameters