7-6
IM DL350-01EN
Scale Value (Scale Value)
Items Whose Scale Values Are Displayed (Display Item)
You can display the upper and lower limits (scale values) of each waveform’s vertical or horizontal axes.
• OFF: Hides the scale values
• ALL: The vertical axis (V Scale) and horizontal axis (Time Scale) are displayed.
• Time Scale: Only the horizontal axis (Time Scale) is displayed.
Horizontal Axis Display Mode (Time Scale Mode)
Set the type of time to display on the horizontal scale.
• Auto: If the horizontal scale (Time/Div) during measurement is greater than 6 min/div, absolute time is used.
Otherwise, relative time is used.
• Relative: Relative time from the start of measurement
• Absolute: Absolute time of measurement (not selectable when the measurement time is 1 second or less)
Trace Label Display (Trace Label)
You can display waveform labels next to the displayed waveforms. If the waveform display is narrow because of
the display format settings, labels may not be displayed.
• ON: Hides labels
• OFF: Shows labels
Level Indicator (Level Indicator)
A level indicator that shows the levels of the waveforms whose displays are turned on appears on the right side
of the waveform display area. It shows the current levels of the sampled data.
Grid (Graticule)
You can set the window grid to one of the following options.
•
: Displays the grid using broken lines
•
: Displays the grid using crosshairs
•
: Displays a frame
Extra Window (Extra Window)
When waveforms and measured values overlap and are difficult to see, you can use the extra window to display
them separately. The extra window appears below the T-Y waveform display window. The following values
appear in the extra window.
• Cursor-measurement values
• Automated measurement values of waveform parameters
• The digital values of each channel (only during roll mode display)
Height of the Extra Window
Set the height of the extra window.
OFF: The extra window is not displayed.
1 to 8: The extra window is set to the selected height.
Auto: The extra window appears automatically when you perform cursor measurements and automated
measurements of waveform parameters.
7 Display