6-8
IM DL350-01EN
AND Trigger (Enhanced)
The instrument triggers on the AND of multiple trigger source conditions. The instrument triggers when all the
specified conditions are met at a single point.
Trigger Source
You can use CH1 to CH6 and Bit 1 to Bit 8 as the trigger sources. Bit 1 to Bit 8 appear as options when
• When a logic input module is installed in a slot.
• When the logic input terminal (CH5 or CH6) of this instrument is in use.
Achievement Condition (Condition)
Set the achievement condition for each trigger source.
H
The signal level must be high.
L
The signal level must be low.
IN
The signal must be within the specified level range.
OUT
The signal must be outside of the specified level range.
−
The signal is not used as a trigger condition.
* IN and OUT are selectable only when the trigger source is an analog signal (CH1 to CH4).
Trigger Level (Level)
Set these items for each trigger source.
When the Achievement Condition Is H or L
Set the level for determining whether the trigger sources are high or low.
When the Achievement Condition Is IN or OUT
An edge is detected when the trigger source enters (IN) or leaves (OUT) the specified level range. You can
specify the level range settings for each analog signal trigger source.
Trigger Level Width (Width)
Set the trigger level width when the edge detection condition is IN or OUT.
Trigger Hysteresis (Hys)
Set these items for each trigger source.
This item is the same as the hysteresis
of the simple trigger.
Trigger Position (Position), Trigger Delay (Delay)
, Trigger delay
6 Triggering