![YOKOGAWA ScopeCorder DL350 User Manual Download Page 113](http://html.mh-extra.com/html/yokogawa/scopecorder-dl350/scopecorder-dl350_user-manual_911346113.webp)
6-6
IM DL350-01EN
Edge On A Trigger (Enhanced)
This is available only in scope mode.
While state condition A is met, the instrument triggers on the OR of multiple trigger source edges.
Edge detection
Trigger
Condition A is being met
Trigger Source
You can use CH1 to CH6 and Bit 1 to Bit 8 as the trigger sources. Bit 1 to Bit 8 appear as options when
• When a logic input module is installed in a slot.
• When the logic input terminal (CH5 or CH6) of this instrument is in use.
State Condition (A State)
To set state condition A, select the states of the trigger sources in relation to the trigger level.
CH1
CH2
CH3
CH4
CH5-
Bit 1
Bit 2
……
Bit 8
CH6-
Bit 1
……
Bit 8
H
L
X
L
X
L
…
X
H
…
L
State Condition
A
Example
H: The signal level must be high.
L: The signal level must be low.
X: The signal is not used as a condition.
Edge Detection Condition (Edge)
Set the condition for detecting the trigger source edge.
An edge is detected when the trigger source changes from a level
below the trigger level to a level above the trigger level (rising).
An edge is detected when the trigger source changes from a level
above the trigger level to a level below the trigger level (falling).
−
The signal is not used as a trigger condition.
Trigger Level (Level) and Trigger Hysteresis (Hys)
Set these items for each trigger source.
These items are the same as the trigger level
of the simple trigger.
State Condition Achievement Condition (Condition)
Select whether the result of comparing the trigger source states to their specified conditions must be true or false
for the state condition to be considered met.
True
The result must be true.
False
The result must be false.
Trigger Position (Position), Trigger Delay (Delay)
, Trigger delay
6 Triggering