10-3
IM 253710-01E
Numerical Computations
10
Setting the period with the cursors
This method can be used to set the period in the normal measurement and
harmonic measurement modes.
For the normal measurement mode carry out steps 3 to 12. For the harmonic
measurement mode, carry out steps 7 and 8.
3.
Press the MEASURE key to display the Measure setting menu.
4.
Press the [Mode] soft key to select [ON].
5.
Press the [Period] soft key to display the period setting menu.
6.
Press the [Cursor] soft key to select cursor setting.
• Setting the start point of the measurement/computation period
(For normal measurement and harmonic measurement modes)
7.
Press the [Start Pos / End Pos] soft key and set the jog shuttle control to [Start
Pos]. For harmonic measurement mode, only the [Start Pos] soft key is
available.
8.
Turn the jog shuttle to set the start position on the waveform.
• Set the end position of the measurement/computation period
(For normal measurement mode only)
9.
Press the [Start Pos / End Pos] soft key to set the jog shuttle control to [End
Pos].
10.
Turn the jog shuttle to set the end position on the waveform.
• Setting the start and end points of the measurement/computation period
simultaneously
(For normal measurement mode only)
11.
Press the [Start Pos / End Pos] soft key to set the jog shuttle control to both
[Start Pos] and [End Pos].
12.
Turn the jog shuttle to set the measurement/computation period. The distance
between them is kept constant.
Go to step 13 on page 10-5.
10.1 Setting the Measurement/Computation Period, Re-computing