PH450 Analyzer 

Proper Calibration Steps 

 

Before any electrodes are placed into initial service and then about once a month 

thereafter, depending on the application, a two point calibration of the electrodes are 
required. A two point calibration corrects for changes in the output of the pH electrode 
(slope) and the millivolt offset (Eo) of the reference electrode, by comparing these 
values to those of known buffer solutions. The EXA Instruments utilize preprogrammed 

buffer tables in the Auto Cal function. These pre-programmed values are based on 
pure N.I.S.T. buffers, which correspond to 6.86 pH, 4.01 pH and 9.18 pH at 25° C. Many 
plants use N.I.S.T. traceable or technical grade buffers, which are pure NBS buffers that 
have been adjusted to read 7.00, 4.00 and 10.00 respectively. If traceable buffers are 
used, and the default values in the preprogrammed Buffer Tables have not been 

changed, an error of .14 pH (7.00 - 6.86 = .14) will occur. To eliminate this situation, either 
pure NBS buffers should be used or the Buffer Tables in the instrument should be 
reprogrammed to match the values of the buffers being used. For further information 
regarding NBS vs NBS Traceable buffers, refer to the Technical Note TNA0917 "A Word 
about...pH Buffers."
 For information regarding the reprogramming of the EXA PH450 refer 

to “Technical Note TNA0905 "PH450 Analyzer Programming Custom Buffer Tables."  
 
When performing a calibration, always start with fresh buffer solutions. You will need two 
buffers (preferably 7.00 and 4.00) and some deionized water to rinse the electrodes. 

Place a sample of the 7.00 buffer solution in the sample cup attached to the flow-
through housing. Take the electrode assembly out of the flow-through housing, clean 
and wipe the probes, then immerse it in the sample cup. 
 
 

1.) Click on the settings icon (wrench) 

 
 
 

 
 
 
 
 

 
 
 
 

 
 

Summary of Contents for PH450

Page 1: ...traceable buffers are used and the default values in the preprogrammed Buffer Tables have not been changed an error of 14 pH 7 00 6 86 14 will occur To eliminate this situation either pure NBS buffers should be used or the Buffer Tables in the instrument should be reprogrammed to match the values of the buffers being used For further information regarding NBS vs NBS Traceable buffers refer to the ...

Page 2: ...ration Wash select it 3 Next selected the diamond next to Automatic zer slope 4 On the following screen in the box at the bottom use the scroll key to select the 6 9 buffer 5 The main display will start flashing and continues to flash until the reading has stabilized for 5 seconds and does not change by more than 02 pH units ...

Page 3: ...7 00 buffer and rinse it thoroughly in deionized water then blot the electrodes to remove any excess water Immerse the electrode assembly in a sample cup of 4 buffer solution 8 On the following screen in the box at the bottom use the scroll key to select the 4 0 buffer 9 The main display will start flashing and continues to flash until the reading has stabilized for 5 seconds and does not change b...

Page 4: ...ore it in memory Press the diamond next to Calibration complete 11 To have the analyzer finish and accept calibration data press the diamond next to Accept Data 12 The analyzer will automatically return you to the original Calibration screen To return to main operating mode press the house icon to return to the main screen ...

Page 5: ... the Main screen press the I icon in the circle The following screen will pop up press the diamond next to NEXT Then the data screen will appear The zero is the AS the slope is the SL and the impedance 2 is the impedance of the reference junction The sensor mV reading is a continuous reading of what the sensor is seeing in normal measureing mode To return to normal measureing mode press the house ...