background image

IM 04P01B01-01E 

6-14

6.13 Performing Calibration Correction (/CC1 

Option)

The values of the measuring input signal are corrected using specified segments, and 
the results are used as measured values.
If you change the input range, set the correction point (measured value) and correction 
values again.

Procedure

1.

  Hold down the 

MENU

 key for 3 seconds to enter Setting mode.

2.

  Press the 

 key or 

SHIFT 

 key to select 

Calibration

 and then press the 

 key.

3.

  Set each item and press the   key.

Use the 

 key or 

SHIFT

 + 

 key to select values.

For the procedure on how to enter values or characters, see section 4.2.
If you press the 

ESC

 key in the middle of the operation, the settings entered up to 

then are cancelled, and the display returns to a higher level menu.

4.

  When the 

Setting complete

 screen appears, do either of the following:

Press the   key to set other channels.
If you are done, press the 

ESC

 key.

5.

  Hold down the 

MENU

 key for 3 seconds to return to Operation mode.

  

Set=Calibration

CH=01

01CH Calibration

Setting complete

Turn On or Off the calibration correction.

Set the first correction point 

(input value).

Select the channel.

(Displays the following when

 Calibration is On)

The new setting takes effect.

Yes: The setting is fixed.

No: The setting becomes invalid.

ESC/?

Calibration=On

1 datums=  0.000

Decision=Yes

Set the first correction value.
Likewise, set the second correction

point and correction value.

..

.

1 revise=  0.000

Summary of Contents for MuR10000

Page 1: ...User s Manual Model 436101 436102 436103 436104 436106 R10000 Recorder IM 04P01B01 01E 8th Edition...

Page 2: ......

Page 3: ...errors please contact your nearest YOKOGAWA dealer as listed on the back cover of this manual Copying or reproducing all or any part of the contents of this manual without the permission of Yokogawa...

Page 4: ...5 1 3x Added Customized menu Section 1 9 Added Modbus register 40301 to 40348 Communication manual BT1 Added Header printout Section 1 4 1 4x EM1 Added SD memory card IM 04P01B01 03E Checking the Ver...

Page 5: ...o the instrument the operator must refer to the explanation in the manual Protective ground terminal AC DC High temperature To avoid injury caused by hot surface do not touch locations where this symb...

Page 6: ...xternal Connection Connect the protective grounding before connecting to the item under measurement or to an external control unit Damage to the Protective Structure Operating the recorder in a manner...

Page 7: ...g this product except those stated in the WARRANTY that is provided separately Use the software on a single PC You must purchase another copy of the software if you are to use the software on another...

Page 8: ...put terminal 4 H3 Non glare door glass H5D Portable type Power cord UL CSA st d 7 H5F Portable type Power cord VDE st d 7 H5R Portable type Power cord AS st d 7 H5J Portable type Power cord BS st d 7...

Page 9: ...code H5F is specified Maximum rated power voltage 250 V A1024WD Provided when optional code H5R is specified Maximum rated power voltage 250 V A1023WD Provided when optional code H5J is specified Max...

Page 10: ...pcs Violet B9902AQ 1 3 pcs Plotter pen Purple B9902AR 1 3 pcs Mounting bracket B9900BX 2 Shunt resistor for the screw terminal standard 415920 1 250 0 1 415921 1 100 0 1 415922 1 10 0 1 Shunt resisto...

Page 11: ...ger on the tab at the lower left of the display and open the display Display Tab on the display Open Open Remove all packing materials Pen Model Hinge Dot Model Hinge CAUTION To protect the hinges do...

Page 12: ...ibes the setup operations for convenient functions such as how to assign tags to channels and how to set message strings that are to be printed 7 Setup Operations for Changing Adding Functions Basic S...

Page 13: ...to prevent such occurrences Calls attentions to actions or conditions that could cause light injury to the user or damage to the instrument or user s data and precautions that can be taken to prevent...

Page 14: ...Blank...

Page 15: ...g 2 11 2 6 Turning ON OFF the Power Switch 2 13 Chapter 3 Names of Parts and Run Operations 3 1 Names of Parts 3 1 3 2 Installing or Replacing the Chart Paper 3 4 3 3 Installing Replacing Felt Pens or...

Page 16: ...Setting the Burnout Detection Function of Thermocouples 7 5 7 4 Setting the RJC Function on Channels Set to TC Input 7 6 7 5 Changing the Channel Recording Color Dot Model 7 8 7 6 Recording by Compens...

Page 17: ...hannel Recording Color Dot Model 9 23 9 14 Changing the Channel Assignments of Recording Pens Pen Model 9 24 9 15 Changing the Type of Report Data Printed in Periodic Printout 9 25 9 16 Setting the Ba...

Page 18: ...Blank...

Page 19: ...currence release and predefined messages Also the recorder settings can be printed Internal Light A light is provided for easier viewing of the recording area of the chart paper Display Measured value...

Page 20: ...used as measured values Also the low cut function input less than 0 is fixed to 0 scale left value can be used Current Input A shunt resistor is attached to the input terminal The current signal is c...

Page 21: ...ls set to scaling or square root computation cannot be designated Scaling The input values are scaled to values in the appropriate unit to be used as measured values 0 V 100 0 C 10 V 300 0 C Measured...

Page 22: ...value B A Scale value Calibration point Correction using absolute values Measured value A Correction value B A1 A2 A3 A4 A5 Related Topics Setting the calibration correction function Sections 7 15 7 2...

Page 23: ...tion can be set for each measurement channel However it does not operate on channels set to ON OFF input Filter Pen Model The filter is a low pass filter The time constant can be set to 2 s 5 s or 10...

Page 24: ...Converter Pen model Select 16 7 ms 60 Hz 20 ms 50 Hz or Auto Dot model Select 16 7 ms 60 Hz 20 ms 50 Hz 100 ms or Auto If Auto is selected the recorder detects the power supply frequency and automati...

Page 25: ...rence High Limit Alarm h An alarm occurs when the difference in the input values of two channels is greater than or equal to the specified value Difference Low Limit Alarm l An alarm occurs when the d...

Page 26: ...lls below the alarm value at time X4 and the alarm is released Note The alarm detection operation is reset when a power failure occurs The operation restarts after the power recovers If the alarm valu...

Page 27: ...rding The alarm occurrence release can be recorded on the chart paper See section 1 4 Alarm Output Relay A1 A2 and A3 Options Contact signals can be generated from alarm output relays when alarms occu...

Page 28: ...I03 I01 and I02 for the A1 option By default the reflash alarm is disabled Alarm Alarm output relay Reflash alarm ON Alarm output relay Reflash alarm OFF Channel 1 Channel 3 Channel 2 Approx 500 ms Ap...

Page 29: ...ion Specifying AND produces no effect If diagnosis output is enabled I01 is set to diagnosis output Specifying AND produces no effect Related Topics Setting the AND operation Section 7 1 Energized De...

Page 30: ...CK Alarm ACK Alarm ACK or or Hold Non hold Activated Released Note If the reflash alarm is enabled I01 to I03 are fixed to non hold operation Specifying Hold produces no effect If diagnosis output is...

Page 31: ...pable of recording the measured values with pens or dots trend recording as well as various other types of information Recording Example on the Pen Model Recording Example on the Dot Model The recordi...

Page 32: ...interval at all times The recording colors in order from channel 01 are purple red green blue brown and black The recording color of each channel can be changed among these six colors For each channel...

Page 33: ...3 pen and 4 pen recorders there are offsets along the time axis phase difference between the pens This offset is corrected when pen offset compensation is used Same time Below is an explanation for th...

Page 34: ...a mark that indicates the position of the date time on the chart paper Time tick cancel mark An exclamation point is printed when the periodic printout time tick was not printed at the correct positio...

Page 35: ...or disabled By default the channel printout is enabled Characters may become discolored or the ribbon may brush against the paper due to the warping of the ink ribbon or the like during a digital pri...

Page 36: ...e Periodic Printout Measured values and other items are printed at the preset interval Printout is not performed when the chart speed is as follows Pen model Less than or equal to 9 mm h or greater th...

Page 37: ...ed A buffer overflow mark is printed when there are messages that cannot be printed because the buffer is full The time printout format can be selected New Chart Speed Printout Printout is not perform...

Page 38: ...t is complete List printout contains Setting Mode settings such as the input range and alarm for each channel Setup List contains Basic Setting Mode settings such as the alarm output relay operation a...

Page 39: ...ons see section 1 10 Function Setup Guide 1 2 3 4 5 6 7 8 9 10 11 12 App Index Printout Example of List on the Dot Model The printout examples may appear differently from the actual printout as a resu...

Page 40: ...om the actual printout as a result of functional improvements Start Printout and End printout When recording starts Start printout or Start printout 2 prints Also when recording stops End printout or...

Page 41: ...neous values are printed out together Up to 5 messages of 35 characters can be entered Messages are printed out in the order in which they are set Message strings set with the standard function are sp...

Page 42: ...when the alarm is released If the alarm indication is set to hold operation the indicator blinks when an alarm occurs When alarm ACK operation is performed the indicator turns ON or OFF MATH This ind...

Page 43: ...ement channel Symbol indicating an alarm Computation channel Channel on which an alarm is occurring Channel on which an alarm is not occurring Channel on which an alarm is not set Date Time and Chart...

Page 44: ...Determines the common logarithm y log10x Exponent EXP Determines the exponent y e x Relational computation LT LE GT GE EQ NE Determines or of two elements and outputs 0 or 1 Logical computation AND OR...

Page 45: ...ts For example if the measured value of channel 01 is 20 mV and the measured value of channel 02 is 20 V the computed result of 01 02 is 40 Recording Computation Channels Trend Recording Trend recordi...

Page 46: ...g using the TLOG timer see appendix 1 Related Topics Setting the periodic printout Sections 7 8 9 6 and 9 15 Setting the printout using the TLOG timer Sections 9 6 and 9 12 Alarms Up to 4 alarms can b...

Page 47: ...hen a CPU failure occurs Therefore relay output is carried out also when the power is turned OFF including a power failure This relay behavior cannot be changed FAIL Relay Output NO C NC Normal NO C N...

Page 48: ...he secondary chart speed while a level signal is applied to the terminal The secondary chart speed is set in advance Message 1 Printout to Message 5 Printout Remote input signal Trigger Prints message...

Page 49: ...on the status of the batch comment switching signal when recording is started stopped remotely When starting stopping by key operation performs Start printout and End printout Starting and Stopping Da...

Page 50: ...the key lock function Section 3 13 Customize Menu The menu can be customized to display only the menus that you use version 1 31 or later Display only the items that you use on the FUNC key menu Displ...

Page 51: ...he date time when switching from the standard time to DST and the date time when switching back from DST to standard time When switching from standard time to DST the clock is set ahead by 1 hour When...

Page 52: ...etting Scaling Range span and scale 5 1 Use Range in Setting mode Unit 5 3 Use Unit in setting mode to set the unit after scaling Square Root Computation Range span and scale 5 1 Use Range in Setting...

Page 53: ...Basic Setting mode to Hold and set Alarm Indicator to Hold Executing the alarm ACK operation 3 12 Use Func Alarm ACK in Operation mode to execute the release of the alarm output Set a hysteresis on t...

Page 54: ...Alarm occurrence release printout Turn printout On Off 7 7 Use Print Alarm in Basic Setting mode to select whether to print the alarm occurrence and release print only the alarm occurrence or not prin...

Page 55: ...ght in Setting mode to set the internal light brightness 6 11 Computation function M1 option Item Description Reference Section Computing equation and recording span Use Math Formula in Setting mode t...

Page 56: ...7 7 Common with measurement channels Periodic printout using TLOG timer dedicated to computation channels Computed value Scale Channel number or tag Recording color Timer number Mar 31 2004 15 50 TAG...

Page 57: ...intout Use Batch in Basic Setting mode to set whether to use the start printout and end printout If Use is selected select a Lot No from 4 digits 6 digits or OFF 7 24 If Use is selected Use Batch Batc...

Page 58: ...Blank...

Page 59: ...iscoloring and deformation Keep electrically charged objects away from the recorder as this may cause malfunction Do not apply volatile chemicals to the door glass display panel keys etc Do not allow...

Page 60: ...l the recorder horizontally However the recorder can be inclined up to 30 degrees backwards for panel mounting Note Condensation may occur if the recorder is moved to another place where both the ambi...

Page 61: ...ccording to the procedure below First attach the two mounting brackets and temporarily fasten the attachment screws Next fix the recorder in place by tightening the attachment screws with the appropri...

Page 62: ...der is an INSTALLATION CATEGORY II product Precautions to Be Taken While Wiring Take the following precautions when wring the input signal cables It is recommended that crimp on lug with insulation sl...

Page 63: ...a nearby fan turns ON or OFF Connecting the input wires in parallel with other devices can cause signal degradation affecting all connected devices If you need to make a parallel connection then Turn...

Page 64: ...ped input terminal Screw input terminal Philips screwdriver Flat blade screwdriver Crimp on lug with insulation sleeves Signal wire Signal wire Note Input signal wires of diameter less than or equal t...

Page 65: ...1 5V input and ON OFF input DC current input Leadwire resistance 10 max wire The resistance of the three wires should be equal Example For a 4 to 20 mA input a shunt resistor of 250 0 1 can be used t...

Page 66: ...o the recorder the terminals of the recorder and or the cable can be damaged In order to prevent tension from being applied directly on the terminals fasten all wiring cables to the rear of the mounti...

Page 67: ...NC A3 NO C NC NO C NC NO C NC NO C NC NO C NC NO C NC A3 R1 F1 I01 I02 I03 I04 I05 I06 I01 I02 I03 I04 I05 I06 NO C NC NO C NC R1 CE FAIL 1 2 3 4 5 C 1 2 3 4 5 C R1 R1 F1 F1 R1 NO C NC NO C NC CE FAIL...

Page 68: ...terminal Remote Control Input Terminals 1 2 3 4 5 C 1 2 3 4 5 C C 1 to 5 Internal circuit Input type Photocoupler isolation Shared common C Allowable input voltage 5 VDC 5V 1 to 5 Remote control input...

Page 69: ...60947 1 3 Connect a fuse in the power supply line Use a fuse approved by CSA for the use in North America or VDE for the use in Europe 2 A to 15 A other than P1 4 A to 15 A P1 option Do not add a swi...

Page 70: ...than P1 P1 24 VDC AC power supply L N L N DC AC24V 3 Close the power supply terminal cover and secure it with the screw The proper torque for tightening the screws is 0 6 N m Portable Type H5x Option...

Page 71: ...urn ON OFF the power switch of the recorder or another device during operation This can have adverse effects on the measured values If nothing is displayed when the power switch is turned ON turn OFF...

Page 72: ...Blank...

Page 73: ...er left and pull to open Mounting hole There is one hole on each of the top bottom left and right panels The hole is covered with a seal Pen model Dot model Display and key panel see the next page The...

Page 74: ...Key Switches the setup item or the value Press this key while holding down the SHIFT key to switch the setup item or the value in reverse order LEFT RIGHT Key Moves the cursor to the right when enteri...

Page 75: ...s where terminals or ports used by options such as alarm output relays and communication interface are installed Ethernet port C7 option Power terminal block The power terminal and protective ground t...

Page 76: ...ation Be sure to replace the chart paper ahead of time When attaching the chart cassette push it in until you hear the left and right stoppers click into place Recordings will be inaccurate if the cha...

Page 77: ...t Chart paper guide 4 Load the chart paper Riffle the chart thoroughly before loading Make sure that the sprocket teeth of the chart drives are properly engaged in the chart paper perforations Make su...

Page 78: ...corder and press the entire chart cassette into the recorder case Push in the chart cassette until it is fixed in place with stoppers Stopper Feeding the Chart Paper 7 Press the FEED key to assure tha...

Page 79: ...t to let the ink come in contact with the eyes mouth or skin If the ink enters the eyes or mouth immediately wash with water If the ink makes contact with your skin immediately wash with soap and wate...

Page 80: ...en pen holder moves near the center position and the Pen exchange End appears Note When the pen moves a line is drawn on the chart paper 4 Replace the pen 5 Press the CH UP key The screen returns to t...

Page 81: ...makes contact with your skin immediately wash with soap and water If you notice any abnormal symptoms immediately consult a doctor 1 Open the door If recording is in progress press the RCD key to stop...

Page 82: ...ult turn the ribbon feeding knob in the direc tion of the arrow to align the ribbon feeding shaft of the cassette with the ribbon feeding shaft hole of the holder Holder tab Hole for the ribbon feed s...

Page 83: ...art feed will not start even when pressing the RCD key if the chart paper is empty or almost out Insert new chart paper before pressing the RCD key Stopping the Recording While recording is in progres...

Page 84: ...puted values the displayed channel is automatically switched in ascending order The switching interval can be set to 1 s 2 s 3 s 4 s or 5 s Measurement channels set to Skip and computation channels se...

Page 85: ...manual printout is executed trend recording is suspended However the recorder continues the measurement and alarm detection in the background When manual printout is complete trend recording resumes I...

Page 86: ...t Start and then press the key The List printout starts The screen returns to the data display screen Print List Start Aborting the List Printout 1 Press the FUNC key 2 Press the key to select Print o...

Page 87: ...tion can be used to prevent unneeded alarm printouts from being executed Procedure 1 Press the FUNC key 2 Press the key to select Buffer clear and then press the key Func Buffer clear 3 Press the key...

Page 88: ...key to select Message and then press the key Func Message 3 Press the key to select message number and then press the key The message printout starts The screen returns to the data display screen Dis...

Page 89: ...red values in periodic printout Procedure Resetting the Report Data 1 Press the FUNC key 2 Press the key to select Periodic and then press the key Func Periodic 3 Press the key to select Reset and the...

Page 90: ...arm is released The alarm output is released when the alarm ACK operation is carried out When an alarm is occurring The alarm output is held even after the alarm ACK operation If the alarm ACK operati...

Page 91: ...asing the Key Lock Note A password is required to release the key lock 1 Press the FUNC key 2 Press the key to select Keylock and then press the key 3 Enter the password for releasing the key lock The...

Page 92: ...Blank...

Page 93: ...nnels and the computing equation unit constant and TLOG setting of computation channels cannot be changed while computation M1 option is in progress The operations that can be carried out in this mode...

Page 94: ...ed from Setting mode Hold down the MENU key for 3 seconds to enter Setting mode Next hold down both the DISP key and the FUNC key for 3 seconds The Basic Setting mode display appears The top and botto...

Page 95: ...make a selection press the key The next screen appears When the Setting complete screen appears the changed item is applied 01 01 Channel Setting complete Using the ESC Key If you press the ESC MENU...

Page 96: ...aracter is to be inserted Press the CHARACTER key or SHIFT CHARACTER key to show Ins DISP and then press the key A space for one character is inserted Enter the character Position to insert the charac...

Page 97: ...3 8 sections 3 9 3 10 section 3 13 section 3 11 section 3 4 Pen exchange End Pen exchange section 3 3 Math key FUNC With the SD memory card EM1 option Data save Data replay SD card Eject menu items a...

Page 98: ...Default Values Pen exchange Pen Model Moves the recording pen to a position that is easily accessible for replacement on the pen model R exchange Dot Model Moves the printer carriage near the center p...

Page 99: ...ight DST DST Start month Start day Start time End month End day End time mm h Tag Periodic are not displayed in the default condition To display these items settings must be changed in Basic Setting m...

Page 100: ...Register User Password Suffix_S P_Adj Pen No Value End End Timeout Duration Keep alive Host Level Suffix primary IP address Subnet mask Primary server Secondary server Gateway Suffix secondary Indica...

Page 101: ...ype Alarm Relay On Off Off Alarm Relay No I01 to I06 I01 Unit CH 01 to 06 01 Unit Unit 6 characters or less Blank Chart Pen Model 82 types pen model 20 mm h Dot Model 1 to 1500 mm h dot model 20 mm h...

Page 102: ...D 0E 0F 0G 0J 0K 0M 0N 0P 0A Math Aux Print out Trend Dot Model On Off On Math Aux Print out Periodic On Off On Math Aux Tag CH 0A 0B 0C 0D 0E 0F 0G 0J 0K 0M 0N 0P 0A Math Aux Tag Tag 7 characters or...

Page 103: ...n Off Off Print CH Tag CH Tag CH Print Channel Dot Model On Off On Print Alarm On1 On2 Off On1 Print Record On On Off Off Print Chart speed On Off On Print Scale On Off On Print Pen color Pen Model On...

Page 104: ...sage5 Priority R_RCD BT1 BatchCMT switch BT1 BatchCMT switch BT1 Data save On Off EM1 None M1 BT1 EM1 option Record On Off Math M1 option Math Timer TLOG Timer No 1 2 1 Math Timer TLOG Mode Off Relati...

Page 105: ...n Calibration CC1 option Calibration CH 01 to 06 01 Calibration Mode Revise Value Abs Value Revise Value Calibration Point 2 to 16 2 Cust menu Cust menu Not Use Not Cust menu Password Numbers and spac...

Page 106: ...04P01B01 01E 4 14 4 3 Menu Structure Settings and List of Default Values Setup Item Pen Dot Selectable Range or Selections Default Value P_Adj Pen No Pen Model 1 2 3 4 1 Store End End Store Abort Sto...

Page 107: ...Range shown on the screen 3 Set each item and press the key Use the key or SHIFT key to select values For the procedure on how to enter values or characters see section 4 2 If you press the ESC key i...

Page 108: ...012 0 F T 200 0 to 400 0 C 328 0 to 752 0 F N 0 0 to 1300 0 C 32 to 2372 F W 0 0 to 2315 0 C 32 to 4199 F L 200 0 to 900 0 C 328 0 to 1652 0 F U 200 0 to 400 0 C 328 0 to 752 0 F WRe 0 0 to 2400 0 C 3...

Page 109: ...R 200 00 01 01 Channel Setting complete Description Scale Left Scale Right and Decimal Place Selectable range mantissa 20000 to 30000 Example The value in the range of 100 00 to 350 00 cannot be speci...

Page 110: ...eft and span right cannot be set to the same value Input Type Range Type Selectable Span Range DC voltage Volt 20 mV 20 00 to 20 00 mV 60 mV 60 00 to 60 00 mV 200 mV 200 0 to 200 0 mV 2 V 2 000 to 2 0...

Page 111: ...e new setting takes effect First channel Last channel ESC Set Range CH 01 01 MOde DI Range Level Span_L 0 Span_R 1 01 01 Channel Setting complete Description Selectable Range Type Span Left and Span R...

Page 112: ...2 V Scale Left Scale Right and Decimal Place Selectable range mantissa 20000 to 30000 Example The value in the range of 100 00 to 350 00 cannot be specified The mantissa of Scale_R is 35000 which exc...

Page 113: ...the right span value Set the decimal place and left scale value Set the right scale value Turn On or Off the low cut function Set the low cut point 0 0 to 5 0 of the recording span 0 1 steps The new...

Page 114: ...electable range of the low cut point 0 0 to 5 0 of the recording span Values below the low cut point is set to 0 scale left value of the recording span Note The displayable printable range of scale va...

Page 115: ...re cancelled and the display returns to a higher level menu 4 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key...

Page 116: ...alarm t Delay low limit alarm Selectable only when the alarm delay function is enabled See section 7 15 Note On channels set to delta computation alarms can be detected on the values illustrated in t...

Page 117: ...alue of the scale is set to 20000 excluding the decimal the alarm value of 20000 is invalid Set a value that is greater than or equal 19999 Difference High Limit Alarm Difference Low Limit Alarm Value...

Page 118: ...ntered up to then are cancelled and the display returns to a higher level menu 4 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done...

Page 119: ...the Setting complete screen appears do either of the following To correct the setting press the key If you are done press the ESC key 5 Hold down the MENU key for 3 seconds to return to Operation mode...

Page 120: ...4 When the Setting complete screen appears do either of the following To correct the setting press the key If you are done press the ESC key 5 Hold down the MENU key for 3 seconds to return to Operati...

Page 121: ...either of the following To correct the setting press the key If you are done press the ESC key 6 Hold down the MENU key for 3 seconds to return to Operation mode Select the trend recording interval m...

Page 122: ...e Auto Chart speed 23 mm h Number of recording channels 6 channels Scan interval 1 s A value that is less than or equal to 900 chart speed and that is the largest integer multiple of the scan interval...

Page 123: ...values or characters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Se...

Page 124: ...ettings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following Press the key to set other channels If you...

Page 125: ...entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following Press the key to set other channels If you are don...

Page 126: ...complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key 6 Hold down the MENU key for 3 seconds to return to Operation mode Set the cha...

Page 127: ...to enter values or characters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 Wh...

Page 128: ...tings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following Press the key to set other channels If you a...

Page 129: ...ocedure on how to enter values or characters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher l...

Page 130: ...e the key or SHIFT key to select values For the procedure on how to enter values or characters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then a...

Page 131: ...and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key 6 Hold down...

Page 132: ...ntered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following To correct the setting press the key If you are don...

Page 133: ...r channels If you are done press the ESC key 5 Hold down the MENU key for 3 seconds to return to Operation mode Select whether to use the bias Set the bias value Set the channel range The new setting...

Page 134: ...ress the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 4 When the Setting complete screen appears do either of the...

Page 135: ...when using revise values A1 A2 A3 A16 A1 B1 A2 B2 A3 B3 A16 B16 A and B are within the measurable range or scaling range of the range type Within the scaling range 5 to 105 of the scale value or 20000...

Page 136: ...to enter values or characters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 4 Wh...

Page 137: ...C Select Detail Selecting Printout On Off for the Batch Name Chart Speed and Date Time Set Batch Mode Start Detail Print Batch name On Chart speed On Batch print Setting complete Select Start End Star...

Page 138: ...Select whether to eject the pen offset compensating data when recording stops Select On or Off POC speed C Speed Select the chart speed when ejecting the pen offset compensating data ESC Select Detail...

Page 139: ...printout 2 is not performed When recording stop detection is performed option F1 recording and printout stop When chart paper is inserted the instrument operates as follows When detected during Start...

Page 140: ...SHIFT key to select MSG format and then press the key 4 Set each item and press the key Use the key or SHIFT key to select values For the procedure on how to enter values or characters see section 4 2...

Page 141: ...rs M D H M Month Day Hour Minute 11 characters M D H M S Month Day Hour Minute Second 14 characters D M H M Day Month Hour Minute 11 characters D M H M S Day Month Hour Minute Second 14 characters D M...

Page 142: ...r Setting mode 2 Press the key or SHIFT key to select Aux and then press the key 3 Press the key or SHIFT key to select DST and then press the key 4 Set each item and press the key Press the key or SH...

Page 143: ...th Oct End day Last Mon End day Last Mon End time 1 00 DST Setting complete Month day of the week and time when the DST ends Month Time Displays the following when the DST is set to Use Explanation DS...

Page 144: ...Select standard time Winter or DST Summer Select On to set the time to switch between standard time and DST Set the time to switch The new setting takes effect Time Summer Summer Off Time 04 07 05 00...

Page 145: ...red when recording is in progress or when computation is in progress on models with the computation function M1 option Procedure Changing the Settings 1 Hold down the MENU key for 3 seconds to enter S...

Page 146: ...cator Nonhold Increase 01 Decrease 01 Hysteresis 0 5 M_Hysteresis 0 0 Alarm Setting complete Applying the Changes and Returning to Operation Mode Press the key or SHIFT key to select End and then pres...

Page 147: ...d Holds the relay output until the alarm ACK operation is executed Indicator Specifies whether to set the alarm indication to hold operation or non hold operation Nonhold Releases the alarm indication...

Page 148: ...do either of the following To correct the setting press the key If you are done press the ESC key Select the integration time of the A D converter The new setting takes effect ESC A D Integrate Setti...

Page 149: ...tion the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following Press the key to set other chann...

Page 150: ...e of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following Press the key to s...

Page 151: ...RJC mode Internal Uses the RJC function on the recorder External Uses an external RJC function If external is selected set the compensation voltage Volt Sets the compensation voltage when using an ex...

Page 152: ...ntered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done...

Page 153: ...the Setting complete screen appears do either of the following To correct the setting press the key If you are done press the ESC key Select On to record by compensating for the pen offset along the t...

Page 154: ...or the procedure on how to enter values or characters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a...

Page 155: ...on CH Tag Channel Prints the channel number Tag Prints the tag Channel Dot Model When On is selected channel number or tag is printed by the trend recording Alarm On1 Prints the alarm information when...

Page 156: ...ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the followi...

Page 157: ...ay returns to a higher level menu 5 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key Select the type of report...

Page 158: ...inimum maximum and average values over the interval SUM Prints the sum value over the interval If SUM is selected set the sum scale MIN Prints the minimum value over the interval MAX Prints the maximu...

Page 159: ...ancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key Set...

Page 160: ...Use the key or SHIFT key to select values For the procedure on how to enter values or characters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to the...

Page 161: ...o the operation of resetting the report data in periodic printout Select Lock to apply the key lock function to the operation of moving the pen to a position where the pen can be easily replaced Keylo...

Page 162: ...K Alarm ACK operation under the FUNC key Math Computation start stop reset operation under the FUNC key M1 option Print out Printout start stop reset operation under the FUNC key Message Message print...

Page 163: ...you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either...

Page 164: ...of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following To correct the sett...

Page 165: ...e middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following To correct...

Page 166: ...lowing To correct the setting press the key If you are done press the ESC key Select the display printout language The new setting takes effect ESC Language Setting complete Basic Language Lang Englis...

Page 167: ...key 4 Press the key with Add function shown on the screen 5 Set each item and press the key Use the key or SHIFT key to select values For the procedure on how to enter values or characters see section...

Page 168: ...ot The 1 5V low cut item does not appear in Setting mode Alarm delay Enables Disables the alarm delay function Use Enables the setting of delay high limit alarm or delay low limit alarm for the alarm...

Page 169: ...er values or characters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 6 When the...

Page 170: ...D H M Month Day Hour Minute M D H M S Month Day Hour Minute Second YMD H M S Month Day Year Hour Minute Second None Does not print the date time RCD On Sets the time printout format of the record star...

Page 171: ...ize and then press the key 4 Press the key to select the settings to be initialized and then press the key Set Initializes the settings in Setting mode Setup Set Initializes the settings in Basic Sett...

Page 172: ...nter values or characters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When th...

Page 173: ...ion function M1 option Math reset Resets the computed result of the computation function M1 option Priority R_RCD Starts stops recording BT1 option BatchCMT switch Start printout and Start printout 2...

Page 174: ...section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen app...

Page 175: ...NU key for 3 seconds to enter Setting mode 2 Hold down the and keys simultaneously for 3 seconds to enter Basic Setting mode 3 Press the key or SHIFT key to select Select menu and then press the key 4...

Page 176: ...display Data replay Select On or Off Select whether to display SD eject Select On or Off EM1 option EM1 option dot model EM1 option Applying the Changes and Returning to Operation Mode Press the key...

Page 177: ...n 5 Set each item and press the key Use the key or SHIFT key to select values For the procedure on how to enter values or characters see section 4 2 If you press the ESC key in the middle of the opera...

Page 178: ...Select On or Off M1 option CC1 option Applying the Changes and Returning to Operation Mode Press the key or SHIFT key to select End and then press the key Press the key to select Store and then press...

Page 179: ...y to select values For the procedure on how to enter values or characters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the...

Page 180: ...mplete Select Cust menu Select Lock Enter the Password Enter Basic Setting mode Select Not The new setting takes effect Menu kind Lock ESC Basic Cust menu Applying the Changes and Returning to Operati...

Page 181: ...enu Procedure 1 Enter a password in the same manner as the procedure for Disabling the Customized Menu press the key and enter Basic Setting mode 2 Change settings in the usual manner 3 The instrument...

Page 182: ...characters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting compl...

Page 183: ...value corrected value measured value Abs Value Sets the correction value using the corrected value Point Selects the number of points that make up the segments including the start and end points in t...

Page 184: ...p to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following To correct the setting the key If you are done press the ES...

Page 185: ...Setting mode Not Does not perform printout of the lot number Dual comment Use You can now select Start printout 2 and End printout 2 in Setting mode s Mode Not There are no Start printout 2 and End p...

Page 186: ...you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either...

Page 187: ...play screen MENU 1 Hold down or 3 s to switch Power ON The panel keys are set to the functions marked above the keys as shown below RCD MENU FEED CH UP FUNC MENU DISP 1 CHARACTER ESC SHIFT Returning f...

Page 188: ...play End DI DO display when the R1 or A option is installed End Display in which different screens can be assigned to the top and bottom sections Channel number Display update rate Top section Channel...

Page 189: ...l switching interval and then press the key 5 When the Setting complete screen appears press the key Hold the MENU 1 key down for 3 seconds to exit from the data display setup screen Select the channe...

Page 190: ...the bottom section Press the key or SHIFT key to select date time and press the key 6 When the Setting complete screen appears press the key Hold the MENU 1 key down for 3 seconds to exit from the dat...

Page 191: ...display comes on Math Start Stopping the Computation 1 Press the FUNC key 2 Press the key to show Math and then press the key 3 Press the key with Stop shown on the screen The computation stops and t...

Page 192: ...p to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press t...

Page 193: ...te a character Clear DISP Delete the entire string Copy DISP Store the entire string in the memory Past DISP Paste the string in the memory Inserting a Character Press the key or SHIFT key to move the...

Page 194: ...ual EQ NE Logical product AND Logical sum and exclusive logical sum OR XOR low order of precedence Data That Can Be Used in Equations In TLOG computation the measured values of measurement channels or...

Page 195: ...e shown below you can determine the accumulation of a specified channel in this case channel 01 and make the result the computed value of computation channel 0A Computing equation of computation chann...

Page 196: ...is not equal to the measured value of channel 3 the computed result is 1 Otherwise the result is 0 02 GE 03 If the measured value of channel 2 is greater than or equal to the measured value of channel...

Page 197: ...e1ORe2 1 e2 0 e1 0 e1ORe2 1 e2 0 XOR Exclusive OR Syntax e1XORe2 Condition If the two data values e1 and e2 are zero and non zero or non zero and zero the computed result is 1 Otherwise it is 0 Expla...

Page 198: ...termines the maximum value of channel e1 TLOG MIN Minimum value Syntax TLOG MIN e1 Condition Determines the minimum value of channel e1 TLOG AVE Average value Syntax TLOG AVE e1 Condition Determines t...

Page 199: ...nter values or characters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When th...

Page 200: ...the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the fol...

Page 201: ...key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following P...

Page 202: ...arm delay time Related Topics Setting the delay time of the delay high limit alarm or delay low limit alarm Section 9 11 Enabling the alarm delay function Section 7 15 Setting the alarm hysteresis Sec...

Page 203: ...T key to select TLOG and then press the key 4 Set each item and press the key Use the key or SHIFT key to select values For the procedure on how to enter values or characters see section 4 2 If you pr...

Page 204: ...utation channels Computation channels Timer mode Absolute time mode Absolute or relative time mode Reset computed TLOG value Yes Yes No selectable Print when computation stops No Yes For the timer 1 a...

Page 205: ...of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 6 When the Setting complete screen appears do either of the following Press the key to set...

Page 206: ...characters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 6 When the Setting compl...

Page 207: ...ss the key Use the key or SHIFT key to select values For the procedure on how to enter values or characters see section 4 2 If you press the ESC key in the middle of the operation the settings entered...

Page 208: ...middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 6 When the Setting complete screen appears do either of the following Press the key...

Page 209: ...the settings entered up to then are cancelled and the display returns to a higher level menu 6 When the Setting complete screen appears do either of the following Press the key to set other channels I...

Page 210: ...reen 5 Set each item and press the key Use the key or SHIFT key to select values For the procedure on how to enter values or characters see section 4 2 If you press the ESC key in the middle of the op...

Page 211: ...interval Absolute Absolute Time Mode After the computation is started the timer times out at the times determined by the reference time on the hour and the interval The Ref Time reference time is set...

Page 212: ...eset operation for sum computation TLOG SUM Example TLOG SUM computation 1 2 3 Reset Reset Reset Reset On Timer timeout Reset Off When reset is ON the sum value is calculated over each interval When s...

Page 213: ...e ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 6 When the Setting complete screen appears do either of the follow...

Page 214: ...le of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 6 When the Setting complete screen appears do either of the following To correct the se...

Page 215: ...then press the key 5 Set each item and press the key Use the key or SHIFT key to select values For the procedure on how to enter values or characters see section 4 2 If you press the ESC key in the mi...

Page 216: ...MAX Prints the maximum value over the interval INST Prints the instantaneous value SUM scale Set this item only when SUM is selected For the definition of the term sum scale see page 7 14 Off Simply...

Page 217: ...he middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 6 When the Setting complete screen appears do either of the following Press the...

Page 218: ...For the procedure on how to enter values or characters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to...

Page 219: ...tation An over value on a measurement channel refers to Over or Over See section 12 4 An over value on a computation channel refers to a value that is less than or equal to 100000000 or greater than o...

Page 220: ...Blank...

Page 221: ...cale limits are equal This is not allowed 24 The lower span limit is greater than the upper span limit 25 The lower scale limit is greater than the upper scale limit 26 Bias cannot be set to the SKIPP...

Page 222: ...er Basic Setting mode during header printout until End printout is finished 172 Action not executable while accessing SD card Cannot be executed while the SD memory card is being accessed 173 Action n...

Page 223: ...ferent If you change a setup item in Basic Setting mode from the Configuration Software load the settings of Basic Setting mode See Loading the Settings of Basic Setting Mode and Setting Mode in Loadi...

Page 224: ...the switch work properly Connect properly Turn ON the power properly Contact your nearest YOKOGAWA dealer Section 2 5 Section 2 6 Supply the power at the specified voltage and frequency Does the input...

Page 225: ...See section 7 2 Use input filter pen model see sections 7 12 and 6 2 moving average dot model see sections 7 11 and section 6 3 Wire correctly see section 2 3 Install the terminal block correctly Fast...

Page 226: ...If an inductive load is connected to an alarm contact output use a surge suppresser on that line Contact your nearest YOKOGAWA dealer Display and other functions do not work properly Is the key lock...

Page 227: ...lines or printed characters clear not blurred For the replacement procedure of felt pens and plotter pens see section 3 3 pen model For the replacement procedure of the ribbon cassette see section 3 4...

Page 228: ...ecommended that the plotter carriage shaft be cleaned once a year Cleaning Procedure Wipe the dust off the plotter carriage shaft using a soft lint free cloth or paper If the dirt does not come off ea...

Page 229: ...e power switch 2 Open the display and key panel section 3 Pull out the internal light LED assembly that is attached to the end of the display and key panel section Display and key panel 4 Remove the L...

Page 230: ...ent Main Specifications Standard temperature stability accuracy 0 05 C For information on purchasing the calibration instruments contact your nearest YOKOGAWA dealer Calibration Procedure 1 Wire the r...

Page 231: ...romotive force table based on 0 C The recorder performs compensation by measuring the temperature at the input terminal and adding the corresponding thermoelectromotive force to the actual output of t...

Page 232: ...for 3 seconds to enter Basic Setting mode 5 Press the key or SHIFT key to select P_Adj and then press the key 6 Press the key to select Zero left edge of the chart paper or Full right edge of the char...

Page 233: ...ote When the Customized Menu is enabled the operation is different When P_Adj in the Customized Menu is On 1 Perform steps 1 4 of Adjusting the Pen Position 2 When the instrument enters Basic Setting...

Page 234: ...paper or Full right edge of the chart paper and then press the key Adjust Hysteresis first followed by Zero and then Full Adjusting the Hysteresis A line is drawn on the chart paper If the line appear...

Page 235: ...the key the changes are discarded and the screen returns to Operation mode Press the ESC key to return to the Basic screen Note When the Customized Menu is enabled the operation is different When P_Ad...

Page 236: ...en 2 km PEN ASSY B9902AM B9902AN B9902AP B9902AQ Red Green Blue Violet At a pen speed of 10 cm s 1 each Plotter pen 60 000 characters PEN ASSY B9902AR When printing continuously 1 Display 5 years DISP...

Page 237: ...200 V range to 4 800 to 5 200 V range TC R1 0 0 to 1760 0 C 32 to 3200 F S1 0 0 to 1760 0 C 32 to 3200 F B1 0 0 to 1820 0 C 32 to 3308 F K1 200 0 to 1370 0 C 328 to 2498 F E1 200 0 to 800 0 C 328 0 t...

Page 238: ...to delta computation the input value of the reference channel where reference channel number channel number set to delta computation Computable input type DC voltage TC and RTD The channel set to delt...

Page 239: ...or each channel common within a channel 1 to 3600 s 1 hour However if the alarm delay time setting is not an integer multiple of the scan interval the alarm delay time is set to a value that is the fi...

Page 240: ...Within the recording span range Trend Recording Dot Model Item Specifications Recording method Trend recording by a 6 color wire dot printer Recording color Channel 1 Purple channel 2 Red channel 3 G...

Page 241: ...r minute month day hour minute second year month day hour minute second and not print Message printout buffer Stores 5 messages waiting to be printed Recording start printout Prints at the start of re...

Page 242: ...odel except offset compensation ON OFF and recording color are not printed For report data Same as the pen model except offset compensation ON OFF and recording color are not printed Message printout...

Page 243: ...1CH digital 4CH bargraph display tag display pen model Status display System display Batch name display BT1 option SD memory card status display EM1 option Lights out Skip Multiple display The follow...

Page 244: ...Computation channels A B C D E F G J K M N P Alarm Alarms are displayed using the symbols below If multiple alarms are occurring simultaneously the alarm with higher precedence is shown Higher preced...

Page 245: ...n is set to hold operation the indicator blinks when an alarm occurs When alarm ACK operation is performed the indicator turns ON stop blinking or OFF Date time and chart speed display Displays the da...

Page 246: ...gits Bar graph scale Displayable values are 19999 to 99999 and a decimal If this range is exceeded the value is displayed as or Channel number 1 digit alarm 1 digit measured value 6 digits Bar graph s...

Page 247: ...rol input R1 option status DO alarm output relay Ax option status Displays the appropriate mark No display Alarm Chart end abcdefg H 1999 9ABCDEF abcdefg H9999999 9ABCDEF abcdefgH 1999 9ABC abcdefgH 1...

Page 248: ...e as the Tag_2CH digital display Channel number 1 digit alarm 1 digit measured value 6 digits Channel number 1 digit alarm 1 digit measured value 6 digits No display Displays the appropriate mark Date...

Page 249: ...stop synchronization Baud rate 1200 2400 4800 9600 19200 or 38400 bps Start bit Fixed to 1 bit Data length 7 bits or 8 bits Stop bit Fixed to 1 bit Parity Odd even or none no parity Received buffer le...

Page 250: ...K01 to K30 Communication input data Digital data via communication can be used in computations other than statistical computation TLOG Number of communication input data points 8 pen model 12 dot mod...

Page 251: ...cy Input Type Measurement Accuracy Recording Accuracy Cu10 GE Cu10 L N Cu10 WEED Cu10 BAILEY Cu10 0 00392 at 20 C Cu10 0 00393 at 20 C 0 4 of rdg 1 0 C Measurement accuracy 0 3 of recording span Cu25...

Page 252: ...tact closed at 200 or less and contact open at 100 k or greater Open collector ON voltage 0 5 V or less sink current 30 mA or more leakage current when OFF 0 25 mA or less Allowable input voltage 5 VD...

Page 253: ...rrected value revise value correction point correction value absolute value correction value within the measurable range and scaling range 5 to 105 or 20000 to 30000 of each input range Setting limits...

Page 254: ...s2 or less Shock Not acceptable Magnetic field 400 A m or less DC and 50 60 Hz Noise Normal mode 50 60 Hz Volt The peak value including the signal must be less than 1 2 times the rated value of the r...

Page 255: ...hat NRTL is included CE EMC EN61326 1 compliance Class A Table 2 For use in industrial locations EN61000 3 2 compliance EN61000 3 3 compliance EN55011 compliance Class A Group 1 Low voltage directive...

Page 256: ...of rdg 1 except R S 0 to 100 C 3 7 C 100 to 300 C 1 5 C B 400 to 600 C 2 C Accuracy not warranted for values less than 400 C 0 1 C K 0 15 of rdg 0 7 C except 0 15 of rdg 1 C for 200 to 100 C E J T 0 1...

Page 257: ...ng accuracy 4 4000 2000 5 2 12 digits 2 When the input is 5 V 100 input A 0 1 5 000 V 2 digits 0 005 V 5 digits 2 digits 7 digits Measuring accuracy 7 4000 2000 2 5 5 6 digits rounded up Recording acc...

Page 258: ...Mounting position With a backward tilt of 30 or less Measurement 0 1 of rdg 1 digit or less Recording Indication fluctuation 0 1 of the recording span or less Vibration When a sinusoidal vibration al...

Page 259: ...hen mounting the recorder to a panel use two brackets top and bottom or left and right Panel Cut Dimensions L mm Side by Side Mounting vertically max 3 units Side by Side Mounting horizontally Units S...

Page 260: ...table Type H5x Option 138 5 5 45 144 5 67 141 5 5 57 12 4 0 49 144 5 67 187 7 36 27 5 1 08 285 11 2 176 6 93 81 5 3 21 100 240V 50 60Hz 40VA MAX Unit mm approx inch Unless otherwise specified toleranc...

Page 261: ...omputation Alarm Time tick Unit Measured value Channel number or tag Date Time Channel data Computation channel Scale Recording color Chart speed Mar 31 2004 15 50 TAG 001 1 000V 2 1 000V 3 H 2 1050UN...

Page 262: ...te time Report end date time Channel data Computation channel Scale Recording color Chart speed Time tick Mar 31 2004 15 00 Apr 01 2004 15 00 TPR TAG 001 1 000V 2 1 000V 3 2 105500E 03UNIT03 4 2 000V...

Page 263: ...as with the instantaneous mode of the pen model Report Mode Sum value Min Max Ave Time tick Unit Range change occurrence mark Measured value Channel number or tag Report start date time Report end dat...

Page 264: ...intout of the pen model Channel data Prints the computation channel numbers A to J or tags computed values and unit Alarm This is the same as with the instantaneous mode of the periodic printout of th...

Page 265: ...d At the next printout time data after the power recovery is used to create and print the statistical calculation data Periodic printout Periodic printout is not performed immediately after power is r...

Page 266: ...mm h_ Mar 31 04 11 30 1 0 001V 2 0 000V 3 0 000V 4 0 000V 5 0 000V 6 0 000V 2 000 2 000 1CH V 20mm h_ The figure above is used to explain the printout contents The actual printout and font are differe...

Page 267: ...ut Items and Interval The next periodic printout does not start until all items of the previous periodic printout are printed If all items cannot be printed in the specified interval the printout inte...

Page 268: ...Blank...

Page 269: ...correction performing of 6 14 calibration instrument 11 4 cassette holder 3 10 changing the settings 4 3 channel auto switching 3 12 channel printout 1 17 channel range 5 10 6 13 channel switching int...

Page 270: ...k 3 19 7 18 keys 3 2 keys to be key locked 7 18 L language 7 22 level 1 31 linear scaling 5 3 List 3 14 loading the chart paper 3 4 logical computation 1 26 9 6 lot number 1 22 lot number setting of 6...

Page 271: ...3 7 replacing the internal light LED 11 3 replacing the plotter pen 3 8 report data 7 14 9 26 report mode App 2 resetting the computed values 9 1 resetting the report data 3 17 resetting the TLOG comp...

Page 272: ...04P01B01 01E U unit 5 12 9 9 unit in computations 1 27 V versions and functions ii VFD 1 24 Volt 5 2 W wiring measuring input 2 7 wiring optional terminal 2 8 wiring power supply 2 11 Z zone recording...

Reviews: