IM 04P01B01-01E
6-14
6.13 Performing Calibration Correction (/CC1
Option)
The values of the measuring input signal are corrected using specified segments, and
the results are used as measured values.
If you change the input range, set the correction point (measured value) and correction
values again.
Procedure
1.
Hold down the
MENU
key for 3 seconds to enter Setting mode.
2.
Press the
key or
SHIFT
+
key to select
Calibration
and then press the
key.
3.
Set each item and press the key.
Use the
key or
SHIFT
+
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the
ESC
key in the middle of the operation, the settings entered up to
then are cancelled, and the display returns to a higher level menu.
4.
When the
Setting complete
screen appears, do either of the following:
Press the key to set other channels.
If you are done, press the
ESC
key.
5.
Hold down the
MENU
key for 3 seconds to return to Operation mode.
Set=Calibration
CH=01
01CH Calibration
Setting complete
Turn On or Off the calibration correction.
Set the first correction point
(input value).
Select the channel.
(Displays the following when
Calibration is On)
The new setting takes effect.
Yes: The setting is fixed.
No: The setting becomes invalid.
ESC/?
Calibration=On
1 datums= 0.000
Decision=Yes
Set the first correction value.
Likewise, set the second correction
point and correction value.
..
.
1 revise= 0.000