5-7
IM 04P01B01-01E
Frequently Used Setup Operations (Setting Mode)
5.1 Setting the Input Range
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
Note
• The displayable/printable range of scale values is –19999 to 30000 excluding the decimal.
• If the leftmost value of the scale is set to –20000 (excluding the decimal), the value when
the low-cut function is applied is displayed as “–Over.”
Square Root Computation
Select SQRT.
Set the channel range.
Select the range type of DC voltage.
20 mV, 60 mV, 200 mV, 2 V, 6 V, 20 V,
or 50 V
Set the left span value.
Set the right span value.
Set the decimal place
and left scale value.
Set the right scale value.
Turn On or Off the low-cut function.
Set the low-cut point.
0.0 to 5.0% of the recording span, 0.1 steps.
The new setting takes effect.
First channel Last channel
ESC/?
Set=Range
CH=01-01
Mode=SQRT
Range=2V
Span_L= -2.000
Span_R= 2.000
Scale_L= 0.00
Scale_R= 200.00
Low-cut=Off
Low-cut point=0.0
01-01 Channel
Setting complete
(When low-cut is On)
(When low-cut is set to Use)
Description
• Input Type in Which Square Root Computation Can Be Performed
Square root computation can be performed when set to DC voltage input.
• Span Left and Span Right
Same as the selectable span range of the DC voltage input.
Make sure that span left is less than span right.