Overview of the GP-IB Interface
51
IM 767361-01E
Overview of the GP-IB Interface
GP-IB Interface Function
Listener
It is possible to enter the same settings via the GP-IB interface that can be
made using the front panel keys.
Measured or computed data, panel setup information, and error codes can be
received.
Talker
Measured or computed data can be output.
Panel setup information and the status byte can be output.
Error codes which have occurred can be output.
GP-IB Interface
Specifications
Electrical and mechanical specifications : Conforms to IEEE St’d 488-1978.
Functional Specifications
: See the table below.
Code
: ISO (ASCII) code
Address setting
: Listener and talker addresses 0 to 30,
or talk-only can be selected.
Remote mode clear
: Remote mode can be cleared by
pressing the
ENTER/LOCAL
on the
front panel. However, this is not
possible if Local Lockout has been
set by the controller.
Functional Specifications
Function
Subset Name
Description
Source handshaking
SH1
Full source handshaking capability
Acceptor handshaking
AH1
Full acceptor handshaking capability
Talker
T5
Basic talker capability, serial polling, untalk on MLA
(My Listen Address), talk-only capability
Listener
L4
Basic listener capability, unlisten on MTA (My Talk
Address), no listen-only capability
Service request
SR1
Full service request capability
Remote local
RL1
Full remote/local capability
Parallel poll
PP0
No parallel polling capability
Device clear
DC1
Full device clear capability
Device trigger
DT1
Full device trigger capability
Controller
C0
No controller function
Turning GP-IB ON/OFF
When starting up using DC power or the battery back, GP-IB is turned OFF.
Turn it ON before use.