![YOKOGAWA mR10000 436101 Manual Download Page 148](http://html.mh-extra.com/html/yokogawa/mr10000-436101/mr10000-436101_manual_911283148.webp)
IM 04P01B01-01E
7-4
7.2 Changing the Integration Time of the A/D
Converter
This section explains the details of setting the integration time of the A/D converter.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the
MENU
key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
3.
Press the
key or
SHIFT
+
key to select
A/D
and then press the key.
4.
Set each item and press the key.
Use the
key or
SHIFT
+
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the
ESC
key in the middle of the operation, the settings entered up to
then are cancelled, and the display returns to a higher level menu.
5.
When the
Setting complete
screen appears, do either of the following:
To correct the setting, press the key.
If you are done, press the
ESC
key.
Select the integration time of the
A/D converter.
The new setting takes effect.
ESC/?
A/D Integrate
Setting complete
Basic=A/D
Integrate=Auto
Applying the Changes and Returning to Operation Mode
Press the
key or
SHIFT
+
key to select
End
and then press the key. Press
the
key to select
Store
and then press the key. The changes are applied, and
the screen returns to Operation mode. If you select
Abort
and press the key, the
changes are discarded, and the screen returns to Operation mode. Press the
ESC
key to
return to the
Basic=
screen.
Explanation
Integrate
Selects the integration time of the A/D converter.
Setting
Integration Time
Scan Interval (Pen Model)
Scan Interval (Dot Model)
50 Hz
20 ms
125 ms
1 s
60 Hz
16.7 ms
125 ms
1 s
Auto
20 ms or 16.7ms
2, 3
125 ms
1 s
100 ms
1
100 ms
–
2.5 s
1: Selectable only on the dot model.
2: Automatically switch by detecting the power supply frequency.
3: The integration time is fixed to 20 ms if “Auto” is selected when using the DC power supply on a
model with the /P1 option.
Summary of Contents for mR10000 436101
Page 2: ......
Page 14: ...Blank ...
Page 18: ...Blank ...
Page 58: ...Blank ...
Page 72: ...Blank ...
Page 92: ...Blank ...
Page 220: ...Blank ...
Page 268: ...Blank ...