<7.MAINTENANCE>
7-7
IM 11G02N01-01E
(3) Carry out the adjustment in the procedure in (4) and subsequent.
● Adjust on the primary side of the optical system so that the values for a to d in 1-1 and 1-2
become as close to 0 as possible within ±100 range.
● Adjust on the secondary side of the optical system so that the values for e to h in 2-1 and 2-2
become as close to 0 as possible within ±100 range.
(4) Operate the optical zero adjustment
Optical zero adjustment knob
Optical
system 1
Optical system 2
F7_3_3E_3.ai
knob to change the value displayed at
a (or e ).
(5) Move the dimmer plate side view to
change the value displayed at b (or
f ).
(6) Move the dimmer plate sidewise to
change the value displayed at c (or
g ).
(7) Move the dimmer plate sidewise to
change the value displayed at d (or
Dimmer plate
Detector set bolt
F7_3_3E_4.ai
h ).
(8) Repeat the procedures in (4) to (7) to
make all the displayed values come
close to 0 as possible within ±100
range.
* Adjust the dimmer plate which is the nearest
to the zero adjustment knob first, and sequentially.
(9) After the optical balance adjustment, mount the top cover of the analyzer unit, then carry out
a moisture interference compensation adjustment, and perform zero and span calibrations.
*
Before moving the dimmer plate, loosen the detector set bolts (just enough to make the plate movable for snug
adjustment).
7.3.4 Moisture interference compensation adjustment method
If the following operation is maladjusted, the measurement may adversely be affected. If you are not
trained for adjustment, do not carry out this operation but contact the distributor or our serviceman.
N
2
or Air
Flow meter
Electrical
Cooling Unit
Water
Bubbler
Drain
F7_3_4E.ai
Proceed to an adjustment if excessively (beyond ± 2%
FS) affected by moisture interference.
After the end of optical balance adjustment, be sure to
carry out moisture interference compensation adjustment.
(1) After warm-up, select the low range, allow dry gas
(N
2
, air) to flow at 0.5 L/min and carry out zero
calibration.
Caution
Summary of Contents for IR400 series
Page 1: ...User s Manual R Model IR400 Infrared Gas Analyzer IM 11G02N01 01E IM 11G02N01 01E 5th Edition ...
Page 7: ...Blank Page ...
Page 11: ...Blank Page ...
Page 33: ...Blank Page ...
Page 35: ...Blank Page ...
Page 83: ...Blank Page ...
Page 93: ...Blank Page ...
Page 97: ...Blank Page ...
Page 117: ...Blank Page ...