<3. INSTALLATION>
3-3
IM 11G02N01-01E
3.2.2 Mounting input/output terminal module
Mount the input/output terminal module on the panel; observing the following method.
(Note) To avoid the effect of noise generated from external units, mount the I/O terminal module
mounting plate on the panel for continuity at the mounting surface and connect the panel
to the same ground as the analyzer main unit.
External dimensions
Mounting dimensions
Mounting method
Terminal
Screwed to panel
Panel plate
To analyzer rear panel
Connection cable
4.5 mounting hole (× 6 places)
M4 screw (× 6 places)
Rectangular hole
302
142142
154
150
150
316
164
p3-03_1.ai
Note) How to ground analyzer main unit and I/O terminal module
To avoid the effect of noises, etc. from external units, it is recommended to ground them by the procedure described below.
* Don’t separate the analyzer and I/O terminal module, and be sure to
ground them together.
Mount the analyzer and I/O terminal module on the
same panel.
Bring I/O terminal module sheet metal and
panel into continuity at ///// portion.
Bring analyzer main unit and panel into continuity
at ///// portion or rail mounting portion.
(No grounding is required at the power terminal).
Ground the panel casing.
F_3.1.2-4.ai
Summary of Contents for IR400 series
Page 1: ...User s Manual R Model IR400 Infrared Gas Analyzer IM 11G02N01 01E IM 11G02N01 01E 5th Edition ...
Page 7: ...Blank Page ...
Page 11: ...Blank Page ...
Page 33: ...Blank Page ...
Page 35: ...Blank Page ...
Page 83: ...Blank Page ...
Page 93: ...Blank Page ...
Page 97: ...Blank Page ...
Page 117: ...Blank Page ...