7-8
IM XL100C-E
7.3
Communication Error Messages
This section gives a list of error codes and messages of the XL100 and their corrective
action. Error responses to communication commands are output in English.
Setting/Basic Setting/Output Communication Command and Setting Data Load Errors
Code
Message
300
Command is too long.
301
Too many number of commands delimited with ‘;’.
302
This command has not been defined.
303
Data request command can not be enumerated with sub-delimiter.
350
Command is not permitted to the current user level.
351
This command cannot be specified in the current mode.
352
The option is not installed.
353
This command cannot be specified in the current setting.
354
This command is not available during sampling or calculating.
Memory Access Errors Using Setting/Basic Setting/Output Communication
Commands
An English error message is returned via the communication interface. It is not shown on
the XL100 display.
Code
Message
360
Output interface must be chosen from Ethernet or RS by using ‘XO’ command.
362
There are no data to send ‘NEXT’ or ‘RESEND’.
363
All data have already been transferred.
Maintenance/Test Communication Command Errors
An English error message is returned via the communication interface. It is not shown on
the XL100 display.
Code
Message
390
Command error.
391
Delimiter error.
392
Parameter error.
393
No permission.
394
No such connection.
395
Use ‘quit’ to close this connection.
396
Failed to disconnect.
Other Communication M
An English error message is returned via the communication interface. It is not shown on
the XL100 display.
Code
Message
400
Input username.
401
Input password.
402
Select username from ‘admin’ or ‘user’.
403
Login incorrect, try again!
404
No more login at the specified level is acceptable.
410
Login successful. (The special user level)
411
Login successful. (The general user level)